LIS
H
The FreeStyle®
Test Strip
Apply blood or control solution to only
one Test Strip Edge
per test.
Important:
Use only one Test Strip Edge per test.
Do not apply blood to both edges. Test strips may
be used only once. Discard used test strips. You can
continue to fill the test strip for up to 60 seconds. Be
sure to reapply the sample to the same Test Strip Edge.
Test Strip Edge
WARNING:
The vial cap is a choking hazard. The cap or vial contains drying
agents that may be harmful if inhaled or swallowed and may cause
skin or eye irritation.
Important Test Strip Information
• The FreeStyle Freedom™ System has an operating range of 40º to
104º F (4º to 40º C).
• Store the test strip package in a cool, dry place between 37º to 86º F
(3º to 30º C).
• Use test strips only within the system operating temperature range.
• Keep away from direct sunlight and heat.
• Store your test strips in their original vial only; never transfer them
to another vial or any other container.
• Never store individual test strips outside the vial.
• After removing a FreeStyle Test Strip from the vial, immediately
replace the vial cap and close tightly.
• With clean, dry hands, you may gently touch the test strip
anywhere when removing it from the vial or inserting it into
the meter.
• Do not use test strips beyond the expiration date. This may cause
inaccurate results.
• Do not bend, cut, or alter FreeStyle Test Strips in any way.
Note
:
For further information, refer to the FreeStyle Test Strip
package insert.