manualshive.com logo in svg
background image

31 

 

CentriMag System Component 

 

No. of Events

 

             Bleeding 

Total 

30 

Total Number of MDRs 

285 

 
The blood leak events listed in Table 14 associated with the cannula were related to failure of the solvent bond between the body of the 
cannula and the integral connector that was used to join the cannulas to the circuit tubing. The designs of the arterial and venous 

cannula were changed in 2015 to use a separate barbed connector to join the cannulas to the circuit tubing.  

 

The motor cable break events listed in Table 14 were the subject of a field action initiated by the applicant in 2018. All previously 
reported complaints with this failure mode were reanalyzed for reportability, resulting in an increase in reportable events related to 

motor cable break failure. Labeling updates and a cable redesign were initiated and have been approved for commercial distribution. 

 

The CentriMag system was the subject of a field action initiated in August 2019 involving reports of motor and pump issues resulting in 
system/alarm(s)/faults from electromagnetic interference (EMI). Root cause was determined to be associated with the calibration of the 

motors manufactured from August 8, 2017 to July 22, 2019. The production test tool software was fixed. All affected CentriMag 
motors are to be returned to an Abbott Medical facility for inspection and recalibration and subsequently returned to the customer. A 
replacement unit will be provided as appropriate. 

 

The overall MDR rate (number of MDR’s/number of pumps distributed)

 

observed for the CentriMag Circulatory Support System is 

consistent with those documented in the published literature for other short-term mechanical circulatory support devices. 
 

Effectiveness Results 

 

The primary analysis to assess effectiveness was based on the 32 evaluable Pivotal FTW Study patients at the 30-day post-device 
removal time point.  The primary study endpoint of survival at 30 days post device removal or to hospital discharge in the pivotal trial 
for failure to wean from cardiopulmonary bypass (FTW from CPB) was 63%, far exceeding the typical rate reported in the literature of 
27%. Key effectiveness outcomes are presented in the tables below. 

 

Table 15. CentriMag Effectiveness in FTW Subjects from CPB Pivotal Study (G030052/S21) – Survival and Primary Endpoint 

Clinical Study 

 

Survival to 
30 Days 
Post-device 

Survival 
To 
Discharge 

Primary 
Endpoint 
Success

21

 

FTW from CPB 
Pivotal Trial 
(G030052/S21) 

32 

 

22/32 
(69%) 

20/32 
(63%) 

20/32 
(63%) 

 
The outcomes summarized in the table above were a function of use of the CentriMag system for durations that range from a minimum 
of 1 day and up to a maximum duration of 90 days. The duration of support observed for the Pivotal Study is summarized below in the 
table below. 

 

Table 16. Duration of CentriMag Support in FTW Subjects from CPB Pivotal Study (G030052/S21) 

Clinical Study 

Indication For Use 

Mean 
Duration 
Of Support 
(days) 

Range 
(days) 

FTW from CPB 
Pivotal Trial 
(IDE G030052) 

FTW 

32 

12.7 

1 - 90 

 
Improvements in hemodynamics were evaluated as a secondary endpoint in the pivotal study for failure to wean from cardiopulmonary 
bypass. As shown in the table below, MAP increased and CVP decreased during support with the CentriMag system during the pivotal 
study for FTW from CPB, based on a comparison of paired values. No paired data points were available for analysis of Left Atrial 
Pressure or Cardiac Index because, given the hemodynamic instability of the patient population at baseline, data for these parameters 
were collected only at the discretion of the investigator.   
 

Table 17. Hemodynamics - FTW from CPB Pivotal Study (G030052/S21) 

Hemodynamic Parameter 

Paired 
Values 
(n) 

Baseline 

Mean Value 
During CentriMag System 
Support 

CVP 

19.2 mmHg 

13.7 mmHg 

MAP 

23 

67.6 mmHg 

75.3 mmHg 

 
Key indicators of end organ function (blood urea nitrogen, creatinine and bilirubin) were tracked during the Pivotal FTW Study of the 
CentriMag system.  Mean blood chemistry values during the first 14 days of support with the device during the pivotal study of failure 
to wean from cardiopulmonary bypass show decreasing trends for BUN and creatinine but increasing levels of bilirubin (Table 18).   

 

Table 18. End Organ Function – FTW from CPB Pivotal Study (G030052/S21)

 

 

Interval 

Mean 

Range 

Blood Urea Nitrogen 

Baseline 

32 

39 

12 – 94 

                                                            

21

 Success: Survival to 30 days post-device removal or to hospital discharge, whichever is longer; or survival to induction to anesthesia for surgery for cardiac transplantation or 

conversion to other long-term mechanical circulatory support system. 

Содержание CentriMag

Страница 1: ...ENTRIMAG CIRCULATORY SUPPORT SYSTEM Abbott Medical Clinical Technical Support Phone number s United States Emergency HeartLine Support USA Tel 1 800 456 1477 Abbott Medical Main Switchboard Tel 1 925...

Страница 2: ...3 Indicates a trademark of the Abbott group of companies Indicates a third party trademark which is property of its respective owner Pat http www abbott com patents 2019 Abbott All Rights Reserved...

Страница 3: ...14 Intraoperative Device Management 15 Guidelines to Prevent Air Entrainment when Initiating Support 15 Patient Care and Management 15 Guidelines to Prevent Air Entrainment during Support 16 CentriMag...

Страница 4: ...5 Safety and Effectiveness and Results 27 Device Malfunctions and Failures 30 Clinical Study Safety and Effectiveness Conclusions 33...

Страница 5: ...ce The CentriMag Circulatory Support System is indicated for use as a right ventricular assist device Humanitarian Device The system when used as a right ventricular assist device is also authorized b...

Страница 6: ...right ventricular assist device TEE transesophageal echocardiography TEG thromboelastography Description The CentriMag system is designed to provide a versatile and effective means for implementing me...

Страница 7: ...t be set high enough to overcome native cardiac pressure in order to prevent retrograde flow Increasing the RPM will increase the flow of blood through the pump CAUTION If an increase in set speed doe...

Страница 8: ...15 20 mmHg are suggestive of right ventricular dysfunction Such a condition is often associated with a decrease in right ventricular contractility and tricuspid insufficiency If adequate LVAD flow can...

Страница 9: ...plies that may be used for this technique 4 CentriMag pumps Drainage cannulas Return cannulas 2nd Generation CentriMag console with motor and flow probe connected Standard 3 8 ID x 3 32 wall tubing Tw...

Страница 10: ...echnique should be performed using aseptic techniques To prime the pump 1 Fill a large sterile basin with three liters of a warm balanced electrolyte solution 2 Slowly submerge one end of the drainage...

Страница 11: ...ailable venous and arterial cannula may be used at the preference of the clinician For central cannulation it is advisable to use cannulas that are wire reinforced to resist kinking The drainage cannu...

Страница 12: ...clamps are removed before the speed is set higher than 1000 RPMs there is a risk of retrograde flow NOTE If other manufacturers cannulas are used follow standard surgical techniques applicable to thos...

Страница 13: ...tions above for left and right sided support See options above for left and right sided support Two cannulation approaches are shown in the figures below The left ventricle is drained via the left ven...

Страница 14: ...E hemodynamic monitoring palpation and direct visualization of the heart will help to determine the volume of blood available for the circuit and the optimal level of flow After the chest is closed th...

Страница 15: ...econds allowing the patient s vascular system to adjust between each RPM change The pump set speed should be gradually increased to the desired flow If flow drops or tubing chatter is observed the set...

Страница 16: ...during insertion line chatter ramping of the flows or flow below the minimum alert Place a stopcock on the open ports of the central line to avoid air being entrained CAUTION Never leave any ports op...

Страница 17: ...Wound sites should be carefully inspected for signs of tissue breakdown or excessive drainage Undue pressure or torque to the surgical site should be avoided in order to minimize trauma with special c...

Страница 18: ...atient during transport Prior to shutting off the power supply and removal of the patient from the transport vehicle briefly unplug the console s power cord to confirm adequate battery charge and cons...

Страница 19: ...d a decreased dependence on inotropic support Initial assessments of ventricular function should be made without increasing inotropic support IABP support or without volume loading of the ventricles E...

Страница 20: ...5 The new tubing connectors are attached using a wet wet connection while taking care to eliminate air at the junction as well as in the circuit Secure these new connections with bands 6 Ensure that...

Страница 21: ...ents suffering from post cardiotomy cardiogenic shock who were unable to be separated from CPB prior to leaving the operating room Clinical Inclusion Criteria Enrollment in the studies was limited to...

Страница 22: ...pport Not applied to this study All possible measures have been attempted to correct low arterial pH arterial blood gas abnormalities electrolytes hypovolemia hypervolemia inadequate cardiac rate dysr...

Страница 23: ...entriMag system removal Adverse events and complications were recorded throughout the duration of CentriMag system support through device removal and until the patient was discharged from the hospital...

Страница 24: ...1 12 Patients N 32 26 12 25 Sex Male 24 75 15 58 8 67 20 80 Female 8 25 11 42 4 33 5 20 Race White 24 75 4 13 4 13 N A13 N A13 19 76 4 16 2 8 African American Other Age mean years SD 58 13 8 59 11 6 5...

Страница 25: ...seline Laboratory Values FTW from CPB G030052 S21 Variable N Mean SD Median Min Max Blood Urea Nitrogen mg dl 32 39 1 21 3 32 12 94 Creatinine mg dl 32 1 8 0 8 1 6 0 9 4 Total Bilirubin mg dl 31 1 8 1...

Страница 26: ...ut not identical across the studies so direct comparisons are not possible However the table shows that general trends in types and incidence of adverse events were similar across all four studies Tab...

Страница 27: ...otension 1 1 3 5 5 19 2 2 17 0 0 0 8 8 8 Hepatic Dysfunction 1 1 3 8 7 27 3 2 17 0 0 0 12 10 11 Renal Failure Dysfunction 8 8 25 3 3 12 1 1 8 12 12 48 24 24 25 Neurologic Dysfunction 0 0 0 8 8 31 2 2...

Страница 28: ...4 4 Other 1219 9 28 0 0 0 0 0 0 2120 10 40 33 19 20 19 Right arm compartment syndrome bronchorrhea and desaturation cardiogenic shock with suspected platelet dysfunction tear in ventricular tissue nea...

Страница 29: ...valuate the safety of the device These data were obtained by analyzing adverse events associated with marketed CentriMag systems which were reported to Abbott Medical between June 1 2014 and June 30 2...

Страница 30: ...he tables below Table 15 CentriMag Effectiveness in FTW Subjects from CPB Pivotal Study G030052 S21 Survival and Primary Endpoint Clinical Study N Survival to 30 Days Post device Survival To Discharge...

Страница 31: ...y Endpoint Clinical Study N Survival to 30 Days Post device Survival to Discharge Primary Endpoint Success Cardiogenic Shock Trial G030052 22 26 11 26 42 Not Defined RVAS Trial G040029 23 12 7 12 58 N...

Страница 32: ...ing acute myocardial infarction or post cardiotomy failure to wean from cardiopulmonary bypass and in patients with acute right ventricular failure from any cause In all studies the adverse event rate...

Отзывы: