25
26
Steps
Do This
Select type of insulin.
9.
If you use this type of
insulin
Then do this
Rapid Acting Insulin Lispro
(Humalog)
Insulin Glulisine (Apidra)
Insulin Aspart (Novolog)
Go to Step 10.
Short acting (Humulin R,
Novolin R)
Scroll to highlight
1.
Type.
Press
2.
Enter button.
Scroll to
3.
Short.
Press
4.
Enter button.
Go to next step.
5.
Select action time.
10.
Scroll to highlight
a.
Action Time, if it is not already highlighted.
Note:
There is a different duration for different types of insulin. The default for
Rapid
is 5h 00m and the default for
Short
is 8h 00m. The action time will be per
default if not changed.
Press
b.
Enter button.
Scroll to desired action time, which ranges between 2h 00m – 8h 00m.
c.
Press
d.
Enter button.
Steps
Do This
Confirm insulin type
11.
and action time is
correct.
Check your settings displayed on the screen.
a.
Notice
b.
Next is highlighted.
Press
c.
Enter button.
Set Inactivity Alarm.
12.
Notice
a.
Time is highlighted.
Press
b.
Enter button.
Scroll to the amount of time when you want the pump to alert you that you
c.
have had NO interaction with the pump. The time ranges between 1h – 24h.
Press
d.
Enter button.
Confirm the time for
13.
inactivity alarm is
correct.
Check the time setting on the
a.
INACTIVITY ALARM screen.
Notice
b.
Next is highlighted.
Press
c.
Enter button.
Section 2: Using the System for the First Time – Quick Setup
2.1 Set Up Your System
Section 2: Using the System for the First Time – Quick Setup
2.1 Set Up Your System
DOC15736_Rev-A-Man.indd 25-26