15
16
Steps
Do This
Check your work.
7.
Caution:
All the signals below must be presented. If any signals below do
•
NOT
occur,
reinsert the battery again or try inserting another fresh battery.
If the problem happens again, do
•
NOT
use the pump. Switch to an alternate
insulin delivery method and call your healthcare professional or Customer
Care.
Notice the
a.
Power up screen will appear.
Listen for one long beep and four quick beeps.
b.
Feel two short vibrations.
c.
Look for the next screen.
d.
If you are
Then do this
Setting up the pump
for the first time
You will see the Set Date/Time screen.
Go to Section 2.2 to Set Up Your System.
Replacing the battery
within 15 minutes
after the pump
has already been
initialized
You will see the
1.
TIME/DATE screen.
Notice
2.
Next is highlighted.
Press
3.
Enter button to return to Home screen.
Reconnect cannula to cannula housing and
4.
then RESTART your basal delivery from your
companion.
2 Using the System for the First Time – Quick Setup
2.1 Set Up Your System
Purpose
To program basic information into your Aviator Pump including time and date, insulin profile,
and
inactivity alarm
, using your Aviator Companion.
When
Do this procedure:
After you have charged the companion’s battery, and installed the pump’s battery.
•
Before you use your companion and pump as a system for the first time.
•
Materials
Your companion fully charged and powered on.
•
Your pump with battery installed.
•
Your companion and pump already paired. This means that your pump is associated with
•
your companion and your pump and companion can only communicate with each other.
Information from your healthcare professional about the type of insulin you will use with
•
your pump and other setting information that you will need as recorded in your Table of
Settings in Section 12 – Appendix A.
Helpful Information
You will perform this task using your companion, and the settings will automatically be set up on your pump as
•
well.
Before entering your settings, be sure to ask your healthcare professional which type of insulin you will be using
•
and what is the length of its action time. Record your personalized settings in the Table of Settings in Section 12
– Appendix A.
When you have more experience with pumping, you can set up your companion with more optional settings in
•
Section 6.
If you are interrupted while doing this setup, the pump will beep to let you know it is waiting for you to finish
•
these steps.
For a detailed list of the screen settings and the order in which they appear, refer to the
•
Freestyle Aviator
Menus in Section 12 – Appendix G.
Section 1: Getting Acquainted
1.4 Install Pump Battery
Section 2: Using the System for the First Time – Quick Setup
2.1 Set Up Your System
DOC15736_Rev-A-Man.indd 15-16
2/9/09 11:29:34 AM