Software Tour
AEROSET
®
Customer Training Guide
1-33
94858-103 — July 2003
Action Area
The following table describes the buttons located in the Action Area.
Press this Button…
To Perform this
Action
Description
Initiate a run or
perform System
configuration
When selected, accesses the
RUN OPTIONS
screen. From this
screen, the Operator selects the types of samples to be run and
initiates the run process. Calibrations and Controls are ordered from
this screen.
SYSTEM CONFIGURATION
is also accessed from
this screen.
PAUSE the System
during a run
When selected, accesses the
RUN OPTIONS
screen. From this
screen the Operator can PAUSE the System which temporarily
stops the movement of the Reagent 1 Probes, Sample Probes,
Carousel Sampler
, and
FastTrack™ Sampler
.
WARNING:
If a diluted sample has been dispensed, the Sample
Arm may move after the status has changed to PAUSE.
Wait 20
seconds before accessing the Carousel Sampler area.
WARNING:
The AEROSET System does not stop the
movement of the Reagent 2 Probes, ICT™ Unit, Mixers, or
Cuvette Washers so the Assay Read Times are not affected.
CAUTION:
The System completes the dispensing and mixing of
any sample and reagent already aspirated.
Confirm the PAUSE
status in the System Status Area and confirm that the R1
indicator on the Local User Interface (LUI) Control Panel is
illuminated before proceeding.
Order and start a
STAT sample
When selected, accesses the screen to order a sample in the
Reserved STAT Position and initiates the processing.
Initiate the
START
UP
Procedure
When selected, allows the Operator to define and initiate the
START UP
Procedures.
Initiate the
SHUTDOWN
Procedure
When selected, allows the Operator to define and initiate the
SHUTDOWN
Procedures.
Scan bar code
labeled Reagent
Cartridges
When selected, initiates a scan of the bar codes on the Reagent
Cartridges in the
Reagent Supply Centers
.
Stop the System
during a run
When selected, completely stops a run (all movement and
processing ceases). Any samples in progress are not completed and
remain pending. When the run is reinitiated by selecting
<RUN>
,
these samples are repipetted.
WARNING: The System attempts to home all robotics
before movement stops. Confirm the READY status in
the System Status Area before proceeding.
!
!
!
!
Содержание Aeroset
Страница 1: ...Abbott Laboratories Abbott Park IL 60064 94858 103 July 2003 AEROSET Customer Training Guide...
Страница 2: ...ii AEROSET Customer Training Guide 94858 103 July 2003 NOTES...
Страница 14: ...xiv AEROSET Customer Training Guide 94858 103 July 2003...
Страница 15: ...AEROSET Customer Training Guide 1 1 94858 103 July 2003 MODULE 1 SYSTEM TOUR MODULE 1 SYSTEM TOUR m...
Страница 56: ...1 42 AEROSET Customer Training Guide 94858 103 July 2003 NOTES MODULE 1 SYSTEM TOUR...
Страница 57: ...AEROSET Customer Training Guide 2 1 94858 103 July 2003 MODULE 2 BASIC RUN MODULE 3 BASIC RUN m...
Страница 90: ...2 34 AEROSET Customer Training Guide 94858 103 July 2003 NOTES MODULE 2 BASIC RUN...
Страница 126: ...3 36 AEROSET Customer Training Guide 94858 103 July 2003 NOTES MODULE 3 DAILY START UP AND CHECKS...
Страница 127: ...AEROSET Customer Training Guide 4 1 94858 103 July 2003 MODULE 4 RUN OPTIONS MODULE 4 RUN OPTIONS m...
Страница 152: ...MODULE 4 RUN OPTIONS 4 26 AEROSET Customer Training Guide 94858 103 July 2003...
Страница 158: ...MODULE 4 RUN OPTIONS 4 32 AEROSET Customer Training Guide 94858 103 July 2003...
Страница 161: ...AEROSET Customer Training Guide 5 1 94858 103 July 2003 MODULE 5 RESULT HANDLING MODULE 5 RESULT HANDLING m...
Страница 201: ...Review Questions AEROSET Customer Training Guide 5 41 94858 103 July 2003 NOTES...
Страница 204: ...MODULE 5 RESULT HANDLING 5 44 AEROSET Customer Training Guide 94858 103 July 2003...
Страница 205: ...AEROSET Customer Training Guide 6 1 94858 103 July 2003 MODULE 6 CALIBRATION MODULE 6 CALIBRATION m...
Страница 244: ...6 40 AEROSET Customer Training Guide 94858 103 July 2003 NOTES MODULE 6 CALIBRATION...
Страница 245: ...AEROSET Customer Training Guide 7 1 94858 103 July 2003 MODULE 7 QUALITY CONTROL MODULE 7 QUALITY CONTROL m...
Страница 287: ...Review Questions AEROSET Customer Training Guide 7 43 94858 103 July 2003...
Страница 290: ...7 46 AEROSET Customer Training Guide 94858 103 July 2003 NOTES MODULE 7 QUALITY CONTROL...
Страница 291: ...AEROSET Customer Training Guide 8 1 94858 103 July 2003 MODULE 8 CONFIGURATION MODULE 8 CONFIGURATION...
Страница 336: ...8 46 AEROSET Customer Training Guide 94858 103 July 2003 NOTES MODULE 8 CONFIGURATION...
Страница 337: ...AEROSET Customer Training Guide 9 1 94858 103 July 2003 MODULE 9 ICT UNIT MODULE 9 ICT UNIT...
Страница 352: ...9 16 AEROSET Customer Training Guide 94858 103 July 2003 NOTES MODULE 9 ICT UNIT...
Страница 353: ...AEROSET Customer Training Guide 10 1 94858 103 July 2003 MODULE 10 MAINTENANCE MODULE 10 MAINTENANCE...
Страница 373: ...As Needed Maintenance AEROSET Customer Training Guide 10 21 94858 103 July 2003 NOTES...
Страница 378: ...10 26 AEROSET Customer Training Guide 94858 103 July 2003 NOTES MODULE 10 MAINTENANCE...
Страница 404: ...11 26 AEROSET Customer Training Guide 94858 103 July 2003 NOTES MODULE 11 COMPONENT REPLACEMENT AND TROUBLESHOOTING...
Страница 405: ...AEROSET Customer Training Guide 12 1 94858 103 July 2003 MODULE 12 ACTIVITIES MODULE 12 ACTIVITIES...
Страница 415: ...AEROSET Customer Training Guide Glossary 1 94858 103 July 2003 GLOSSARY GLOSSARY GLOSSARY...
Страница 437: ...AEROSET Customer Training Guide Appendix 1 94858 103 July 2003 APPENDIX HAZARDS APPENDIXHAZARDS m...
Страница 450: ...Appendix 14 AEROSET Customer Training Guide 94858 103 July 2003 NOTES HAZARDS...
Страница 451: ...AEROSET Customer Training Guide Index 1 94858 103 July 2003 INDEX...
Страница 456: ...Index 6 AEROSET Customer Training Guide 94858 103 July 2003 NOTES Index...