background image

SS ZMT

AN-85

ORDERING INFORMATION

ZIRCONIA ZMT OXYGEN ANALYZER SYSTEM

Select one character or set of characters from each category and specify complete catalog number per sample below.

Code No.

Description

BASE NUMBER - 1st thru 3rd characters

ZMT

ZMT Oxygen Analyzer

PROBE TYPE - 4th character

1

ZGP2 (Probe to be ordered separately)

2

ZFG (Probe to be ordered separately)

3

Other

REFERENCE AIR SUPPLY - 5th character

0

None

1

Pump+Air Gauge (available only with Probe ZGP)

2

Regulator (ZFG) (available only with Probe ZFG)

3

Pump (ZFG) (available only with Probe ZFG)

4

Regulator (available only with Probe ZGP)

PROBE TEMPERATURE CONTROL - 6th character

0

None (ZGP2 option only)

1

Temperature Control (ZFG option only)

1st FUEL OPTION - 7th and 8th
characters

_ _

See Table A.

2nd FUEL OPTION - 9th and 10th
characters

_ _

See Table A.

PROGRAM - 11th character

0

Basic

1

Efficiency

2

Humidity in air

3

Auto Cal. - Basic (see Note 1)

4

Auto Cal. - Efficiency (see Note 1)

UNUSED - 12th character

0

None

OUTPUT MODULE 1 - 13th character

0

None

4

relay

OUTPUT MODULE 2 - 14th character

0

None

1

One relay

4

relay

OUTPUT MODULE 3 - 15th character

0

None

1

One relay

2

Two relays (Required with 11

th

 character 3 or 4)

4

Relay - Not available with Probe ZGP (see Note 2)

MAINS VOLTAGE - 16th character

1

110V  50/60 Hz

2

230V  50/60 Hz

ZMT2310000014001   SAMPLE CATALOG NUMBER

Notes:

1. 11

th

 character Auto Cal. Selection requires the selection of two relays in Output Module 3.

2. May be selected when using efficiency program (11

th

 character) to provide second analog output.

Table A

Code

Option

Code

Option

00

No Fuel Specification

09

Distillate Oil

01

Natural Gas

10

No. 4 Fuel Oil

02

Propane

11

No. 5 Fuel Oil

03

Butane

12

No. 6 Fuel Oil

04

Medium Oil

13

Coal (general)

05

Heavy Oil

14

Bituminous Coal

06

General Fuel Oil

15

Steam Coal

07

Naphtha

16

Anthracite

08

Kerosene

17

Coke

SUNSTAR传感与控制 http://www.sensor-ic.com/ TEL:0755-83376549 FAX:0755-83376182E-MAIL: [email protected]

SUNSTAR自动化 http://www.sensor-ic.com/ TEL: 0755-83376489 FAX:0755-83376182 E-MAIL: [email protected]

Содержание ZMT Zirconia

Страница 1: ...for maximum flexibility Two levels of program security to reduce accidental changes High levels of accuracy resolution and RFI immunity CEGB approved ZMT Zirconia Oxygen Analyzer Systems ABB Instrumentation SUNSTAR传感与控制 http www sensor ic com TEL 0755 83376549 FAX 0755 83376182E MAIL szss20 163 com SUNSTAR自动化 http www sensor ic com TEL 0755 83376489 FAX 0755 83376182 E MAIL szss20 163 com ...

Страница 2: ...the range of 600 C to 1400 C 1112 to 2552 F when used with the ZGP2 probe The temperature of the probe heater is automatically controlled by the ZMT when used with the ZFG2 zirconia probe Probe reference air supply is available either through pump units or by use of an air regulator operated from the customer s instrument air line Construction and Operation The ZMT zirconia analyzer is housed in a...

Страница 3: ...ha 16 Anthracite 08 Kerosene 17 Coke Basic Type Ref Probe Type Ref Air Supply See Note 2 Probe Temp Control Fuel Option See Note 1 Efficiency Unused Output Module 1 Output Module 2 Output Module 3 Mains Voltage Code Digit Nos 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 ZMT 1 ZGP2 0 1 4 None Pump Air Gauge Regulator 0 None 1st 00 01 02 03 04 05 06 07 08 09 10 11 12 13 14 15 16 17 2nd 00 01 02 03 04 05 0...

Страница 4: ...r 230V 15 50 60Hz Power consumption 150VA Insulation mains to earth line to ground 2kV r m s Outputs and Set Points Analog outputs isolated 0 to 10mA 0 to 20mA or 4 to 20mA up to three max into 1kΩ max load Output Modules 1 and 2 programmable O2 any range 5 min span within the range of 0 to 25 O2 i e dual ranging possible Temperature any range 200 C 392 F min span within the ranges Cell temp 0 to ...

Страница 5: ...ases are in contact with the outer electrode producing a potential proportional to O2 concentration The Probe The zirconia probes constructed from 316 stainless steel house a ceramic dust filter and flame trap the detector cell cell heater and thermocouple Wiring between the electronics unit and the probe can be carried in a single 6m 20ft flexible conduit which is PVC coated for IP65 NEMA4 rated ...

Страница 6: ... certified test gas Flue Temperature 20 to 600 C 68 to 1112 F Pressure Suitable for all normal positive or negative flue pressures Dimensions See dimension details diagram Probe Fixing Flange or 21 2 in NPT screwed fitting ZFG2 0 4m Probe Insertion 0 4 1 0 1 5 or 2 0m standard Specials up to 4 0m max Flange Options 0 4m ZFG2 Standard 6 0 0 4 thick x 101 0 1 dia 6 holes 7 3 dia equispaced on 80 0 0...

Страница 7: ...erature compensation to be made A gas calibration port is provided to enable the probe output to be checked using test gas mixtures without removal from the process Reference air is required for accurate operation Principle of Operation The detector cell is constructed from stabilized zirconia with inner and outer electrodes attached and is specific to oxygen Air is supplied to the inner reference...

Страница 8: ...ted NiCr NiAl or Pt Pt 13 Rh Thermocouple Pt Pt 13 Rh Pt 2 BS4937 Type R NiCr NiAl Pt 4 BS4937 Type K Insertion Length 556mm 21 9in or 953mm 37 5in Weight 2 5kg 5 5lb nett 600mm 24in probe 2 8kg 6 16lb nett 1000mm 39in probe Code Digits Insertion Length Thermocouple Sheath Mounting Connector Head 1 2 3 4 6 7 8 9 10 ZGP2 Zirconia Oxygen Probe 1 2 9 600mm 24in 1000mm 39in Special 1 2 9 Pt Pt13 Rh BS...

Страница 9: ...Efficiency see Note 1 UNUSED 12th character 0 None OUTPUT MODULE 1 13th character 0 None 4 Analog relay OUTPUT MODULE 2 14th character 0 None 1 One relay 4 Analog relay OUTPUT MODULE 3 15th character 0 None 1 One relay 2 Two relays Required with 11 th character 3 or 4 4 Analog Relay Not available with Probe ZGP see Note 2 MAINS VOLTAGE 16th character 1 110V 50 60 Hz 2 230V 50 60 Hz ZMT231000001400...

Страница 10: ...Instrumentation Inc ABB Kent Taylor SpA St Neots Cambs P O Box 20550 22016 Lenno England PE19 3EU Rochester New York 14602 0550 Como Italy Tel 01480 475321 Tel 716 292 6050 Tel 0344 58111 Fax 01480 217948 Fax 716 273 6207 Fax 0344 56278 SUNSTAR传感与控制 http www sensor ic com TEL 0755 83376549 FAX 0755 83376182E MAIL szss20 163 com SUNSTAR自动化 http www sensor ic com TEL 0755 83376489 FAX 0755 83376182 ...

Отзывы: