TB82 Conductivity Series ATEX Product Instruction Manual
ABB
10
TB82COND-ATEX-EN-A
ELECTRICAL CONNECTIONS - Power
The TB82 Transmitter is powered by DC voltage. See the
appropriate power requirements in the Specification
Section for the particular type of TB82 Transmitter in use
(Analog, HART, PROFIBUS or F
OUNDATION
fieldbus (FF)).
Power is connected to the POWER/OUTPUT area on
TB1, terminals 1 and 2 (see
Figure 7
).
The terminal blocks located in the rear of the transmitter
accept wire sizes from 12 to 24 AWG. Pin-style
terminations should be used for all connections. The
terminal block label identifies all electrical connections and
should be followed when wiring the transmitter.
Normal grounding practice is to terminate all grounds at
the control room side, in which case the field side of the
screen should be adequately protected to avoid contact
with metallic objects. The transmitter case should be
grounded. Ground connections are provided internally (in
the terminal compartment) and externally.
For IS systems the grounding should be at the safety
barrier earth connection. For bus-powered systems the
grounding of the screen should be close to the power
supply unit. The specific noise immunity and emitted
interference are only guaranteed when bus screening is
fully effective (e.g., ensuring that screening is maintained
through any existing junction boxes.) Appropriate equi-
potential bonding must be provided to avoid differences in
potential among the individual plant components.
To ensure fault-free communication on fieldbus (FF or PA)
installations, the bus must be properly terminated at both
ends. Only approved bus terminators must be used for
intrinsically safe circuits.
REAR COVER
REMOVED
EXTERNAL
GROUND SCREW
T01088E
PANEL GASKET
PANEL
9.5
0.38
MAXIMUM PANEL
THICKNESS
EPTH
NEL D
PANEL
CUTOUT
MINIMUM PA
= 144.8 MM (5.70 IN.)
11.9
0.47
135.4
+1.3
–0.8
5.33
+0.05
–0.03
135.4
+1.3
–0.8
5.33
+0.05
–0.03
11.9
0.47
LOCK ASHER (4)
REAR VIEW
PANEL MOUNTING
BRACKET (4)
3/8-IN.
W
3/8-16 x 1/2-IN.
HEX SCREW (4)
PANEL MOUNTING
SCREWS (4)
Figure 6 – Panel Mounting Detail
Warning
.
Use this equipment only in those classes of
hazardous locations listed on the nameplate.
Uses in other hazardous locations can lead to
unsafe conditions that can injure personnel and
damage equipment.
Figure 7 - Power Connections
SENSOR CONNECTIONS
TB2
POWER/OUTPUT
CONNECTIONS
POWER
TEST
1
2
3
4
5
6
7
8
+1
- 2
+3
- 4
SIGNAL
AMETER
TERMINALS
TB1
Содержание TB82EC
Страница 7: ...ABB TB82 Conductivity Series ATEX Product Instruction Manual TB82COND ATEX EN A 7 Figure 2d ...
Страница 8: ...TB82 Conductivity Series ATEX Product Instruction Manual ABB 8 TB82COND ATEX EN A Figure 2d ...
Страница 35: ...ABB Manuel d instructions Produits ATEX Série Conductivité TB82 TB82COND ATEX FR A 35 ...
Страница 36: ...Manuel d instructions Produits ATEX Série Conductivité TB82 ABB TB82COND ATEX FR A 36 ...
Страница 63: ...ABB Bedienungsanleitung zu den ATEX konformen Leitfähigkeits Analysatoren der Serie TB82 TB82COND ATEX DE A 63 ...
Страница 64: ...Bedienungsanleitung zu den ATEX konformen Leitfähigkeits Analysatoren der Serie TB82 ABB 64 TB82COND ATEX DE A ...
Страница 89: ...ABB Manuale di istruzioni di ATEX serie conducibilità TB82 TB82COND ATEX IT A 89 ...
Страница 90: ...Manuale di istruzioni di ATEX serie conducibilità TB82 ABB 90 TB82COND ATEX IT A ...
Страница 91: ...ABB Manuale di istruzioni di ATEX serie conducibilità TB82 TB82COND ATEX IT A 91 ...
Страница 92: ...Manuale di istruzioni di ATEX serie conducibilità TB82 ABB 92 TB82COND ATEX IT A ...
Страница 119: ...ABB Manual de instrucciones del transmisor de conductividad TB82 ATEX TB82COND ATEX SP A 119 ...
Страница 120: ...Manual de instrucciones del transmisor de conductividad TB82 ATEX ABB 120 TB82COND ATEX SP A ...
Страница 141: ...ABB Manual de instrucciones del transmisor de conductividad TB82 ATEX TB82COND ATEX SP A 141 ...