SCC-U
UTILITY UNIT | OI/SCC-U-EN REV. B
19
9
Dismounting and disposal
Dismounting
WARNING
Risk of injury due to process conditions.
The process conditions, for example high pressures and
temperatures, toxic and aggressive measuring media, can
give rise to hazards when working on the device.
• Before working on the device, make sure that the process
conditions do not pose any hazards.
• If necessary, wear suited personal protective equipment
when working on the device.
• Depressurize and empty the device / piping, allow to cool
and purge if necessary.
NOTICE
Damage to the analyzer system!
Damage to the analyzer system and its components due to
improper decommissioning.
• Before the utility unit is shut down it must be purged to
prevent the accumulation of condensate and deposits.
• Make sure the utility unit is free of residual moisture that
can freeze if low temperatures are encountered during
storage and transport.
Shutting down the utility unit
1. Switch off the power supply of the external devices using the
corresponding circuit breakers.
2. Switch off the power supply using the main switch.
3. Shut off the sample gas supply to the utility unit.
4. Remove the gas lines from the connections on the utility unit.
5. Thoroughly purge the utility unit gas paths with an inert gas.
6. Fully tighten the gas connections.
7. Disconnect the electrical leads.
• Use suited tools to disassemble the device, taking the weight
of the device into consideration.
• If the device is to be used at another location, the device
should preferably be packaged in its original packing so that
it cannot be damaged.
• Observe the notices in
Disposal
Note
Products that are marked with the adjacent symbol
may
not
be disposed of as unsorted municipal waste
(domestic waste).
They should be disposed of through separate
collection of electric and electronic devices.
This product and its packaging are manufactured from materials
that can be recycled by specialist recycling companies.
Bear the following points in mind when disposing of them:
• As of 8/15/2018, this product will be under the open
scope of the WEEE Directive 2012/19/EU and relevant
national laws (for example, ElektroG - Electrical
Equipment Act - in Germany).
• The product must be supplied to a specialist recycling
company. Do not use municipal waste collection points.
These may be used for privately used products only in
accordance with WEEE Directive 2012/19/EU.
• If there is no possibility to dispose of the old equipment
properly, our Service can take care of its pick-up and
disposal for a fee.
Содержание SCC-U
Страница 22: ...22 SCC U UTILITY UNIT OI SCC U EN REV B 12 Appendix Technical drawings Dimensions ...
Страница 23: ...SCC U UTILITY UNIT OI SCC U EN REV B 23 ...
Страница 24: ...24 SCC U UTILITY UNIT OI SCC U EN REV B 12 Appendix Technical drawings Internal construction ...
Страница 25: ...SCC U UTILITY UNIT OI SCC U EN REV B 25 Pneumatic diagram ...
Страница 26: ...26 SCC U UTILITY UNIT OI SCC U EN REV B 12 Appendix Technical drawings Wiring diagram Page 1 of 4 ...
Страница 27: ...SCC U UTILITY UNIT OI SCC U EN REV B 27 Page 2 of 4 ...
Страница 28: ...28 SCC U UTILITY UNIT OI SCC U EN REV B 12 Appendix Technical drawings Page 3 of 4 ...
Страница 29: ...SCC U UTILITY UNIT OI SCC U EN REV B 29 Page 4 of 4 ...
Страница 30: ...30 SCC U UTILITY UNIT OI SCC U EN REV B Notes ...
Страница 31: ...SCC U UTILITY UNIT OI SCC U EN REV B 31 Notes ...