SCC-U
UTILITY UNIT | OI/SCC-U-EN REV. B
11
Gas connections
1
Outlet to sample gas cooler (standard)
2
Sample gas inlet (standard)
3
Outlet to gas sampling probe (option)
4
Test gas inlet 1 (option)
5
Test gas inlet 2 (option)
6
Test gas inlet 3 (option)
Figure 7: Gas connections on the rear panel
Sample gas inlet conditions
The sample gas must be non-corrosive, non-flammable and non-
toxic. The utility unit has one gas path.
Sample gas pressure
p
abs
= 70 to 105 kPa (0.7 to 1.05 bar)
Test gas pressure
p
abs
≤
200 kPa (2 bar)
Note
If gas pressures > 250 kPa are expected, a suitable external
pressure reducer must be connected upstream.
Sample gas flow rate
• 70 l/h with SCC-S
• 100 l/h with SCC-C and SCC-F
Sample gas inlet temperature
10 to 50 °C
Note
The relevant data in the gas analyzer data sheets must be
considered as well for the sample handling components.
Gas connections
The gas connections are PVDF fittings for DN 4 / 6 mm tube.
Connecting the gas lines
• Use gas lines made from material that is suited to the
measuring task.
• Connect the gas lines to the gas inlet and outlet fittings on
the rear panel.
F00233
1
4
2
5
6
7
Содержание SCC-U
Страница 22: ...22 SCC U UTILITY UNIT OI SCC U EN REV B 12 Appendix Technical drawings Dimensions ...
Страница 23: ...SCC U UTILITY UNIT OI SCC U EN REV B 23 ...
Страница 24: ...24 SCC U UTILITY UNIT OI SCC U EN REV B 12 Appendix Technical drawings Internal construction ...
Страница 25: ...SCC U UTILITY UNIT OI SCC U EN REV B 25 Pneumatic diagram ...
Страница 26: ...26 SCC U UTILITY UNIT OI SCC U EN REV B 12 Appendix Technical drawings Wiring diagram Page 1 of 4 ...
Страница 27: ...SCC U UTILITY UNIT OI SCC U EN REV B 27 Page 2 of 4 ...
Страница 28: ...28 SCC U UTILITY UNIT OI SCC U EN REV B 12 Appendix Technical drawings Page 3 of 4 ...
Страница 29: ...SCC U UTILITY UNIT OI SCC U EN REV B 29 Page 4 of 4 ...
Страница 30: ...30 SCC U UTILITY UNIT OI SCC U EN REV B Notes ...
Страница 31: ...SCC U UTILITY UNIT OI SCC U EN REV B 31 Notes ...