68
FEM610 |
PROCESSMASTER | OI/FEM610-EN REV. A
8 . . .Diagnosis / error messages
. . .Overview
Error no . / Range
Error text
Current output
Digital output
Pulse output
LCD display
Error maskable?
M038.010 /
Electronics
Sensor memory
defective. Mem. or
connect. defective.
Replace memory.
Current value - no
change.
No response.
Current value - no
change.
Current value -
no change.
Menu "...Group
Masking".
M037.015 /
Electronics
NV chips defect on
Motherboard.
Defective MB.
Replace MB. Call
Service.
M032.023 / Config.
Curr.Out 31 / 32 not
calibrated. Call
Service.
M028.008 / Config.
Display value is <
1600 h at Qmax.
Change mass Unit or
vol. Unit for Totalizer.
M026.005 / Config.
Maintenance interval
is reached. Perform
maintenance.
M024.009 / Config.
Device not
calibrated.
Call Service.
High Alarm or Low
Alarm, depending on
parameter "Curr.Out at
Alarm".
Current value - no
change.
Current value - no
change.
Current value -
no change.
Menu "...Group
Masking".