NOTICE – Equipment damage.

 Take great care to avoid any leaks associated with this 

detector, as leaks can cause a variety of problems. See Troubleshooting in section 

3

. 

2.4

 

Scheduled maintenance 

There are no parts within the detector power supply or electrometer which require scheduled 

maintenance. 

2.5

 

Recommended spare parts 

 

Part Number 

Description 

Quantity 

852A038-1 

Power Supply DBDID 

885A005-1 

Detector Cell Assembly DBDID 

851K006-1 

FID/DBDID Amplifier Kit  

(Order 851K006-1 Amp Kit to build the standard 

851A094-11412 per included 851J003 bulletin.) 

 

Contact your local ABB sales and service representative for specific instructions for ordering spare 

parts. See the last page of this instruction for contact information. Always include the information 

listed in Equipment Identifications and Configuration Identification in your request. 
When ordering parts for replacement, use the list of parts included in the Engineering Data Package 

that was provided with the analyzer to insure the correct version of each part. 

Figure 2-6: DBDID drawing 

 

 

3

 

Troubleshooting 

This section describes general troubleshooting procedures, issues, and remedies. 
Before proceeding with further troubleshooting procedures, first verify: 

 

The system has power. 

 

All switches and gas settings are correct. 

 

Column flow rates and pressure settings are correct. 

 

Using a VOM meter, verify the voltage between chassis ground and the test points on the 

bottom of the power supply board are as follows: 

 

+10 volts DC Bias voltage 

 

+2.5 volts DC Feedback voltage 

 

+2.5 volts DC Plasma voltage 

Содержание PGC5000 Series

Страница 1: ...ABB MEASUREMENT ANALYTICS 892J009MNAE PGC5000 Gas Chromatograph Dielectric Barrier Discharge Ionization Detector DBDID SERVICE INSTRUCTION Measurement made easy...

Страница 2: ...Symptom Spikes 11 4 Specifications 12 4 1 Physical 12 4 2 Power 12 4 3 Modes of operation 12 4 4 Detector sensitivity 12 4 5 Gas consumption 12 5 Drawings 12 List of Figures Figure 1 1 Dielectric Bar...

Страница 3: ...d safety regulations concerning the construction and operation of electrical installations The regulation on technical working materials safety guidelines for tools The regulations and recommendations...

Страница 4: ...DBDID uses a high voltage alternating current discharge to create a plasma The system operates at a high voltage and a low current which is applied to the glass reactor tube at the round metal electr...

Страница 5: ...high voltage alternating current applied to a gas as it flows through a dielectric material The gas is typically helium or argon and the dielectric material is a quartz or Pyrex tube Two electrodes a...

Страница 6: ...injury High voltage present Make sure that power to the PGC5000 has been disconnected before removing the detector To remove the detector from the PGC5000 perform the following steps Disable the isoth...

Страница 7: ...t the coax cable item 1 in Figure 2 3 from the DBDID amplifier assembly Figure 2 3 DBDID amplifier assembly 1 On the DBDID power supply board unplug the high voltage power connector J2 item 3 the bias...

Страница 8: ...g the correct flow 1 Set flows by measuring the flow at the cell vent tube Refer to the data pack for correct flows The detector should automatically light when power is applied To verify remove the f...

Страница 9: ...ee the last page of this instruction for contact information Always include the information listed in Equipment Identifications and Configuration Identification in your request When ordering parts for...

Страница 10: ...a tube in a dark environment There should be a pink or almost white glow in the plasma tube If the tube is blue and the signal is low there is nitrogen in the plasma Verify that the correct reaction g...

Страница 11: ...system 3 6 Symptom Baseline drift Possible Cause Action s Poor pressure control Pressure controls not providing a consistent pulse free flow rate to the detector can cause detector drift Monitor the p...

Страница 12: ...ed Argon ionization mode AID In this mode the detector is not sensitive to fixed gases and methane but is sensitive to a wide range of organic and inorganic volatile components It is capable of ionizi...

Страница 13: ...to optional purge system 2 Reaction gas inlet plasma gas 3 Purge gas vent Detector assembly purge not used in most applications Flameproof breather must remain open or connected to optional purge syst...

Страница 14: ...14 DBDID 892J009MNAE Figure 5 3 Diagram flow PGC5000 air and carrier...

Страница 15: ...892J009MNAE DBDID 15 Figure 5 4 PGC5000 oven electronics internal connections diagram sheet 1...

Страница 16: ...16 DBDID 892J009MNAE Figure 5 5 PGC5000 oven electronics internal connections diagram sheet 2...

Страница 17: ...echnical changes or modify the contents of this document without prior notice With regard to purchase orders the agreed particulars shall prevail ABB does not accept any responsibility whatsoever for...

Отзывы: