7
2
INSTALLATION
Fig. 2.5 Relay Contact Protection
NC C NO
External
DC Supply
+
–
Relay Contacts
Load
Diode
NC C NO
External
AC Supply
L
N
Relay Contacts
C
R
Load
A – AC Applications
B – DC Applications
Fig. 4.1 Nitrate Electrode
3
SETTING UP
Note. Before proceeding any further, ensure that all switches
are set to OFF on the right hand side of the electronics unit –
see Fig. 2.3.
a) Ensure that all external electrical and plumbing connections
have been made correctly.
b) Fill reagent and standard solution bottles and connect them to
the monitor. (See Section 8.1 for details of these solutions.)
c) Fit the probe according to the instructions in Section 8.2.6.
d) Connect the electrical supply and switch on.
Note. The temperature controlled block requires up to half an
hour to reach the normal control temperature. During this
time, ‘Temp. Control Error’ is indicated on the display. Any
calibrations are prevented by the microprocessor during this
time.
e) Verify that there is an adequate supply of sample to the
monitor constant head unit.
f)
Fit the pump platen on the peristaltic pumps (see Section
8.2.7) and switch the pumps on with the switch on the side of
the monitor. Ensure that the peristaltic pumps rotate, and
check that sample and reagents are being drawn into the
monitor by observing the progress of any small bubbles
present in the inlet tubes.
g) Run the monitor for at least one hour to allow the temperature
to stabilize, solutions to be pumped into the system and to
purge the air from the pipework. Check for any leaks around
the pipe connections and rectify as necessary.
h) If the monitor exhibits good stability, i.e. ±2% of reading, carry
out a calibration – see Programming Page.
i)
Check the condition of the sample filter and replace it if
necessary. Ensure that new filters are fitted correctly by taking
note of the flow directions indicated on the filter bodies.
Output
Module 1
Link for the Current
Output required
Output Module 2
(if fitted)
Temperature Input
Module
Input
Module
Protection
Plate
Captive
Screws
0 to 1mA
0 to 10mA
0 to 20mA
4 to 20mA
1
8
1
8
1
8 1
8