![ABB Nitrate 8236 Скачать руководство пользователя страница 4](http://html1.mh-extra.com/html/abb/nitrate-8236/nitrate-8236_instruction-manual_3045463004.webp)
2
1.1
Description
The Model 8236 Nitrate Monitor is a microprocessor based
analyser using an ABB nitrate ion-selective probe. This
equipment is used for environmental water monitoring.
1.2
Training
Due to the specialised nature of the above instrument, it is
recommended that, where the end users personnel have had no
previous experience of maintaining this equipment, training be
provided by this Company.
Such training is available via the local Company in the UK, or
Overseas Agent elsewhere and may be carried out either on the
users premises or at the factory.
1.3
Location and Function of Main Components –
Fig. 1.1
Monitoring of any sample by ion-selective electrodes usually
requires certain conditioning to be performed on the sample to
allow accurate, trouble-free measurement. Such conditioning
involves regulating flow, controlling temperature and adjusting
chemical composition prior to presentation of the sample to the
point of measurement. This monitor carries out these
adjustments for each parameter in simple and easily understood
steps. The sample is presented from the user's pipework to a
constant head unit and any excess is allowed to overflow. This
allows sample, under controlled pressure conditions, to be
presented to a multi-channel peristaltic pump which proportions
sample and reagent solutions through the monitor to the mixing
point. The resulting combined solution is then temperature
controlled to remove the effects of sample and ambient
temperature variations. The conditioned solution is then delivered
to the flowcell, under constantly flowing conditions, where the ion
measurement takes place. This measurement is carried out
using an ion-selective sensing electrode.
The sensor generates a millivolt output which is proportional to
the ion concentration. This output is processed by the
microprocessor-based electronics section to calculate the actual
sample concentration.
To maintain optimum measurement accuracy it is necessary to
introduce standard solutions of known concentration for
calibration purposes. The monitor utilises solenoid valves to
introduce these standard solutions automatically, at
predetermined intervals, under the control of the microprocessor.
The electronics section consists of the main Microprocessor Unit
situated at the top left and a User Junction Box at the top right
behind the hinged case.
Fig. 1.1 Location of Instrument Components
1
INTRODUCTION
Nitrate 8236
Calibration
Solution
Containers
Constant Head Unit
(behind case)
with Sample Inlet and
Drain Connections
Peristaltic
Pump
Case Lock
Heated
Sensor
Flow Cell
ISE,
Reference Electrode and
Temperature Sensor
Cable Entry Glands
Case Lock
Microprocessor
Unit
Hinged Case
Solenoid Valves
(behind case)
User Junction Box
(behind case)
Hinged Door
Push-to-Release
Latch