2

INF12/113–EN Rev. C

| Navigator 550 | Low level dissolved oxygen wet-section | Spares replacement procedures

3 Safety

4 Replacement procedures

Warning.

These procedures must be carried out by a trained technician.

CHEMICAL

Ensure personal protective equipment (PPE) such as

gloves

and

eye

protection

are

worn

during

any

maintenance.

Observe all health and safety procedures for handling
chemicals.

To familiarize yourself with handling precautions, dangers
and emergency procedures, always review the Material
Safety

Data

Sheets

prior

to

handling

containers,

reservoirs and delivery systems that contain chemical
reagents and standards.

Take care if cleaning any spillages and observe all relevant
safety instructions. Wipe up any spillages using clean
water.

Perform general cleaning of the wet-section using a damp
cloth only – mild detergent can be used as a cleaning aid.
Do not use Acetone or any organic solvents.

ELECTRICAL

Isolate all high voltage supplies to the transmitter before
performing replacement procedures.

The wet-section is vulnerable to electrostatic damage.
Wear an anti-static strap or dismantle the wet-section on
an anti-static workbench.

Ensure all electrical connections are kept dry at all times.

GENERAL

Shut off the external sample supply to the wet-section
before performing replacement procedures. When a
procedure is complete, restore power to the transmitter,
sample to the wet-section at the correct flow rate and, if
necessary, calibrate the wet-section – refer to the
wet-section Operating Instructions (OI/ADS550-EN) for
calibration instructions.

Dispose of the old components in accordance with the
guidelines

contained

in

the

Operating

instructions

(OI/ADS550-EN).

Fig. 4.1 Low level dissolved oxygen wet-section – replacement procedures

Modbus cable assembly –

see Section 4.5, page 5

PCB housing O-rings,

PCB cover and PCB

see Section 4.1, page 3

Tundish assembly –

see Section 4.2, page 3

Flowmeter assembly –

see Section 4.3, page 4

Drain valve assembly

and drain valve cable –

see Section 4.4, page 4

Содержание Navigator 550

Страница 1: ...t made easy 1 Introduction These procedures must be carried out by a trained technician Tools required Pozidrive screwdriver Slot head terminal screwdriver Anti static strap 2 For more information Fur...

Страница 2: ...olvents ELECTRICAL Isolate all high voltage supplies to the transmitter before performing replacement procedures The wet section is vulnerable to electrostatic damage Wear an anti static strap or dism...

Страница 3: ...the new small PCB housing seal C between the PCB cover B and PCB housing F a small amount of silicone sealant or grease can be used to help retain the seal in place 11 Refit the PCB cover B using the...

Страница 4: ...art numbers Drain valve assembly AW502 056 Drain valve cable assembly AW502 085 Referring to Fig 4 6 1 Depress the manual override button A on the drain valve B to drain the wet section 2 Disconnect t...

Страница 5: ...cable gland A at the wet section PCB housing B and disconnect the Modbus cable D from the terminal block connections C marked black red screen green and white 3 Withdraw the Modbus cable D from the PC...

Страница 6: ...Notes 6 INF12 113 EN Rev C Navigator 550 Low level dissolved oxygen wet section Spares replacement procedures Notes...

Страница 7: ...Notes Navigator 550 Low level dissolved oxygen wet section Spares replacement procedures INF12 113 EN Rev C 7 Notes...

Страница 8: ...n this document and in the subject matter and illustrations contained therein Any reproduction disclosure to third parties or utilization of its contents in whole or in parts is forbidden without prio...

Отзывы: