MagMaster
Electromagnetic flowmeter
3 Programming & Configuration
4
IM/MM2WM–EN Rev. A
3 Programming & Configuration
3.1 Startup
The first time power is applied the HMI screen displays the following message:
Press 'No'. This screen does not appear on subsequent power up.
3.2 Accessing the Programming Menus
Refer to WaterMaster Programming Guide (IM/WMP) and login at Advanced security level.
3.3 Programming Menu – Sensor Setup
The 'Sensor Setup' programming menu enables the WaterMaster electronics to be 'Retro-Fitted' to
MagMaster MF/F sensors.
For retrofit applications an extra menu (Sensor Setup) appears at the end of the menu tables; it is not
present on factory-calibrated WaterMaster flowmeters. Note that after completing configuration of the
retrofit installation, this menu disappears.
If this menu is not present and you are unable to configure the necessary calibration factors below, contact
ABB Service.
The 'Sensor Setup' menu (while visible) enables the setting of items such as the [Sensor Type], [Sensor
Size] and calibration [Span Ss] and [Zero Sz] at 'Advanced' level access. These items are normally editable
only at 'Service' level and are hidden after setup.
[Sensor Setup]
[Sensor Type]
Type of Flow Sensor, set to WM Full Bore:
[Probe]*
[DE4]*
[Process]*
[DE2]*
[Hygienic]*
[Process OIML]*
[WM Reduced Throat]*
[Hygienic OIML]*
[WM Full Bore]
[No Sensor]*
*Do not use these settings
Note.
When [Cal. Status] = Not Complete,
alarm appears on the display and 'S110.034
Sensor setup not complete' alarm appears in the active alarms list.
No
Yes
Blank Sensor Detected
System Startup
Copy Image From Tx?
Menu
Exit
Select
Sensor Setup
SP033 Adam Street Wynnum SPS - WaterMaster - Electromagnetic Flowmeter -
OIML R49 Class 1 450mm - User Guide
Q-Pulse Id: VM486
Active: 08/04/2016
Page 6 of 12