background image

C

I/

LL

T

10

0

-E

N

 

R

ev

. A

 

12

.2

0

17

We reserve the right to make technical 
changes or modify the contents of this 
document without prior notice. With  
regard to purchase orders, the agreed 
particulars shall prevail. ABB does not 
accept any responsibility whatsoever for  
potential errors or possible lack of 
information in this document.

We reserve all rights in this document and 
in the subject matter and illustrations 
contained therein. Any reproduction, 
disclosure to third parties or utilization of 
its contents – in whole or in parts – is 
forbidden without prior written consent  
of ABB.  
© ABB, 2017

ABB Inc. 
Measurement & Analytics 
Level Measurement Products 
3400, Rue Pierre-Ardouin 
Québec (Québec) G1P 0B2 
Canada
Tel.: +1 800 858-3847 (North America) 
Tel.: +1 418 877-2944 (other countries) 
Fax: +1 418 877-2834
email: [email protected] 

abb.com/measurement

5  General model

Sensor  

offset

Vessel 

height  

(URP)

Level

Ullage

Vessel full  

(20 mA)

(URV)

Vessel empty  

(4 mA)

(LRV)

6  Specification

Environmental conditions

Operating temperature

–40 to 60°C (–40 to 140°F),  
up to 280°C (535°F)  
with cooling tube

Altitude

 Up to 2000 m (6561 ft.)

Relative humidity

 0 to 100%

Electrical equipment

 Class III

Pollution degree

 4 (IP66 / IP67 / Type 4X)

Overvoltage

 Category 1

Output

Analog

 4 to 20 mA, NAMUR compliant

Digital

 HART 7 (multi-variable output) 

Communication

 Local HMI, EDD / DTM, handheld 

Power supply

Powered from the loop

 4 to 20 mA, 15.5 to 42 V DC

Heated lens option

 24 V DC (3W)

 

Отзывы: