2.4.1   

Setting ALARM to 3.6mA (Low Alarm)

 

 

Enter the ALARM SETUP MODE by applying power to the unit while Magnet is at position X. 

 

Slide magnet to position Z to set the unit to 3.6 mA / Low Alarm; this step is verified by checking that 
the output is now at 3.6 mA 

 

2.4.2   

 Setting ALARM to 21mA (High Alarm) 

default

 

 

Enter the ALARM SETUP MODE by applying power to the unit while Magnet is at position X. 

 

Slide magnet to position S to set the unit to 21.0 mA / High Alarm; this step is verified by checking 
that the output is now at 3.6 mA 

 

Note: 

Once in ALARM SETUP MODE, you can toggle the High & Low alarm settings as many times as 

needed. After 5 seconds of no switch activity, the unit will time out and resume normal operation using 
the new alarm setting. 
 

2.5 

4 mA / 20 mA DAC Trim 

If required, the unit’s 4 ma and 20 ma outputs can be trimmed as follows: 
 

2.5.1   

 Setting 4.0 mA DAC Trim

 

 

Enter the 4 mA DAC TRIM MODE by applying power to the unit while Magnet is at position Z. 

 

Magnet at position Z will decrease the output at - 0.01 mA intervals 

 

Moving the magnet to position S will increase the output at + 0.01 mA intervals 

 

Continue to Increase/Decrease the output until the output is at 4.00 mA 

 

The unit will time out after 5 seconds of no switch activity and resume normal operation using the 
new 4.0 mA trimmed value. 

 

2.5.2   

Setting 20.0 mA DAC Trim

 

 

Enter the 20 mA DAC TRIM MODE by applying power to the unit while Magnet is at position S. 

 

Magnet at position S will increase the output at + 0.01 mA intervals 

 

Moving the magnet to position Z will decrease the output at - 0.01 mA intervals 

 

Continue to Increase/Decrease the output until the output is at 20.00 mA 

 

The unit will time out after 5 seconds of no switch activity and resume normal operation using the new 20.0 mA trimmed 
value.

 

 

 

 

Содержание K-TEK AccuTrak AT500

Страница 1: ...AT500 0200 1b 10 2010 DCN0528 ...

Страница 2: ...tting 4 2 4 1 Setting ALARM to 3 6 mA Low Alarm 5 2 4 2 Setting ALARM to 21 mA High Alarm 5 2 5 4 mA 20 mA DAC Trim 5 2 5 1 Setting 4 0 mA DAC Trim 5 2 5 2 Setting 20 0 mA DAC Trim 5 3 TROUBLESHOOTING INFORMATION 7 3 1 Proper Powerup 7 3 2 Output Stability 7 3 3 Conversion from AT500 to AT600 or AT600 to AT500 7 4 APPENDIX A 8 4 1 Installation Drawings 8 5 APPENDIX B 10 5 1 FMEDA Certificate 10 6 ...

Страница 3: ...ff of the magnet assembly within the float riding on the sensor tube 2 INSTALLATION AND STARTUP WARNING When fitted with a compression fitting the sensor tube is shipped with a set of TEFLON ferrules and a set of STAINLESS STEEL ferrules in a separate bag The TEFLON ferrules are only in tended for low pressure applications for higher operating pressures and or permanent installation replace the TE...

Страница 4: ...d 2 3 Reversing Action If required transmitter output can be reversed by following these steps Note this reverses the 4 20 mA output Bring the level to 50 or 10 Enter the calibration mode by placing Magnet at position X for 2 seconds Slide magnet from position X to position Z to temporarily set the ZERO 4 00mA setpoint this step is verified by checking that the output is now at 4 00 mA Bring the l...

Страница 5: ...equired the unit s 4 ma and 20 ma outputs can be trimmed as follows 2 5 1 Setting 4 0 mA DAC Trim Enter the 4 mA DAC TRIM MODE by applying power to the unit while Magnet is at position Z Magnet at position Z will decrease the output at 0 01 mA intervals Moving the magnet to position S will increase the output at 0 01 mA intervals Continue to Increase Decrease the output until the output is at 4 00...

Страница 6: ...NT DAC 01 mA SW Z DECREMENT DAC 01 mA POWER UP SW Z ENTER 4 0 mA DAC TRIM MODE SW S INCREMENT DAC 01 mA SW Z DECREMENT DAC 01 mA POWER UP SW X ENTER ALARM MODE SETUP SW S SET 21 mA ALARM SW Z SET 3 6 mA ALARM POWER UP SW X 2 sec ENTER CALIBRATION MODE SW S SET 20 mA 100 POINT SW Z SET 4 mA 0 POINT NORMAL OPERATION Magnetic switchpoints are located on lines S X Z at points 90 degrees from the condu...

Страница 7: ...ted a loss of signal no float detected problem with the configuration or a malfunc tion will cause the output to be set to the alarm condition of 3 6 mA 3 2 Output Stability If the current output occasionally jumps up or down but not to an alarm condition This can be an indication of noise or grounding problem Field wiring should be run with an individually shielded cable Insure that the transmitt...

Страница 8: ...AT500 0200 1b 10 2010 DCN0528 8 4 APPENDIX A 4 1 Installation Drawings ...

Страница 9: ...AT500 0200 1b 10 2010 DCN0528 9 4 1 Installation Drawings continued ...

Страница 10: ...AT500 0200 1b 10 2010 DCN0528 10 5 APPENDIX B 5 1 ...

Страница 11: ...er alterations misuse abuse improper or inadequate adjustments applications or servicing of the product K TEK s warranty does not include onsite repair or services Field service rates can be supplied on request If a product is believed to be defective the original purchaser shall notify K TEK and request a Returned Material Authorization before returning the material to K TEK with transportation p...

Страница 12: ...AT500 0200 1b 10 2010 DCN0528 12 6 2 RMA Form ...

Страница 13: ...AT500 0200 1b 10 2010 DCN0528 13 6 3 EU Declaration of Conformity ...

Страница 14: ...AT500 0200 1b 10 2010 DCN0528 14 ...

Отзывы: