14
Assembly of Probe 2
180
650
1
2
α
1
Mounting for the inlet filter
2
Inlet filter (inner filter) with internal heating rod
3
Gas sampling tube
4
Wall tube with inlet flange
5
Sample gas outlet and test gas/purge gas inlet G 1/4
L1
Fitting length
Lx
Length of the gas sampling tube (approx. 400 mm)
x
Minimum distance of the flange of probe 2 from the wall
The following table shows the recommended mounting angle and the minimum
distance x
min
of the assembly flange from the wall depending on the mounting
angle of the probe 2:
Mounting angle
α
10° 15° 20° 25° 30° 35°
x
min
/mm
133 138 143 147 151 153
Step Action
1
Align the gas sampling probe, so that the opening of the inlet filter
(inner filter) is not directly in the process gas flow.
2
Attach probe 2 to the flange of the wall tube with the enclosed
screws M12 x 70.
3
Connect the gas lines to the sample gas outlet and the test gas/purge
gas inlet.