Messages and tests
42 - EN
FSM4000
D184B141U02
Warning code and
ID letter
Prio-
rity
Description
Possible cause
Corrective action
Warning: 10
Reverse Q
8
If the flow direction is forward,
the warning is displayed in case
of reverse flow.
Reverse flow
direction, possibly
defective check valve.
Prevent reverse flow or
change flow direction to
forward/reverse.
Warning: 11
2)
Poll. Adr. > 0
12
The HART unit address was
changed to an address other
than zero. The current output
remains frozen at 4 mA.
Address 1 ... 15
selected. The current
output thereby
remains set at 4 mA.
Select 0 as address, if
the current output
should be 4 ... 20 mA.
Warning. 12a
2) 3)
Simulation Iout
13
Test the current output. Check
the value displayed at the
connection terminals using a
digital voltmeter or the process
instrumentation.
Function test, enter
current output directly
in mA for simulation.
Exit the functional test
of current output.
Warning. 12b
2) 3)
Simulat. Pulse
14
Test the pulse output. Check the
value of output frequency with a
counter.
Simulation pulse
output on.
Exit the functional test
of the pulse output.
Warning. 13
2) 3)
Auto. Adjust
15
Adjustment started for system
zero point adjustment via the
contact input for internal
adjustment (for ABB Service
only).
The external contact
is closed
For ABB Service
personnel only.
The external contact is
open again.
For ABB Service
personnel only.
Warning. 14
Hold MV
16
Warning results only when the
"Noise reduction" function is
activated.
Measuring signal
affected by powerful
interference signal.
Set noise reduction to
lower level or
deactivate. If
necessary, contact ABB
Service.
Warning F
Warn. Diagnosis
17
Diagnosis or wiring warning
Min-Max
over/undershoot for
diagnostic values.
Wiring error for
electrode circuit. Coil
short-circuit to ground
Retrieve details of the
warning message
under the Status
submenu in the
Diagnostic warning
register. Adjust the limit
value, if necessary.
Retrieve details of
warning message
under the Status
submenu in the Wiring
warning register.
1)
Warning is displayed for 30 seconds.
2)
Transmitter must be equipped with HART communication protocol.
3)
Warning is displayed only if the function is called by an external HART protocol command.