| Endura AZ40 | Oxygen and carbon monoxide equivalent (COe) analyzer

4 MODBUS communications

This section describes the connection of serial data cables 
between the master (host computer) and slave (Endura AZ40 
transmitter) on a MODBUS serial link.

4.1 Host computer serial communications

The serial interface option module operates using the MODBUS 
Remote Terminal Unit (RTU) Master / Slave protocol. An 
appropriate RS422 / 485 communications driver must be fitted 
to the host (master) computer. It is strongly recommended that 
the interface has galvanic isolation to protect the computer 
from lightning damage and to increase signal immunity to noise 
pick-up.

IMPORTANT (NOTE)

MODBUS menus are enabled only if a MODBUS 
communications module is fitted.

Menu

Comment

Default

MODBUS

 

Device Address

Set a device-specific address (from 1 to 247) to identify the transmitter on a 
MODBUS link. 

RS485 Setup

 

 

Mode

Select the MODBUS serial communication serial link type:

2 Wire

4 Wire

 or 

Off

 

4 Wire

Baud Rate

Select a communication transfer rate:

1200

2400

4800

9600

19200

38400

 or 

115200

 baud.

19200

Parity

Select the parity bit (transmission error-checking) condition:

No Parity

Odd Parity

 or 

Even Parity

No Parity

Tx Delay

Set a delay to the response from the transmitter in milliseconds. 
Maximum delay 100 ms.

50 ms

IMPORTANT (NOTE)

The MODBUS option provides the following facilities:
— Standard RS422/485 communications.

— MODBUS RTU protocol – the transmitter is 

designed to operate as a Remote Terminal Unit 
(RTU) slave when communicating with a master 
(host) system.

— 500V DC isolation from external connections to the 

instrument.

— 2- or 4-wire communication (selected at the 

RS485 

Setup / Mode

 parameter).

— 1200, 2400, 4800, 9600, 19200, 38400, or 115200 

baud transmission rate.

— Parity-checking (odd, even or none). 

DANGER – Serious damage to health / risk to life

Refer to the Endura AZ40 Operating instruction 

(OI/AZ40-EN) before making electrical connections.

Содержание Endura AZ40

Страница 1: ...s together with MODBUS reference tables for the Endura AZ40 analyzer For more information Further publications for the Endura AZ40 analyzer are available for free download from www abb com analytical...

Страница 2: ...f specification alarms 10 A 1 6 Transmitter and sensor diagnostic maintenance alarms 10 A 1 7 Media card diagnostics 11 A 1 8 Relay state diagnostics 11 A 1 9 Alarm state diagnostics 11 A 2 Digital ou...

Страница 3: ...re injury WARNING Bodily injury High temperature This symbol in conjunction with the signal word WARNING indicates a potentially dangerous situation Failure to observe this safety information may resu...

Страница 4: ...tested half mask conforming to EN140 or equivalent with a level 3 particulate filter conforming to EN 143 or equivalent Disposable protective coveralls in accordance with Type 5 ISO 13982 1 2004 or e...

Страница 5: ...btained from the Company together with servicing and spares information WARNING Bodily injury The transmitter weighs 7 6 kg 17 lb and must be mounted in accordance with the information supplied in the...

Страница 6: ...c disposal systems after 12 August 2005 To conform to European local and national regulations EU Directive 2002 96 EC European electrical equipment users must now return old or end of life equipment t...

Страница 7: ...below the Select prompt The Communication page appropriate to the communication module fitted is displayed MODBUS refer to Section 4 page 8 IMPORTANT NOTE Communication level menus are enabled only if...

Страница 8: ...transmitter on a MODBUS link RS485 Setup Mode Select the MODBUS serial communication serial link type 2 Wire 4 Wire or Off 4 Wire Baud Rate Select a communication transfer rate 1200 2400 4800 9600 19...

Страница 9: ...tions Make connections to the MODBUS serial board as shown in Fig 4 1 On systems with multiple slaves make connections to two or four wire link configurations in parallel as shown in Fig 4 2 When conn...

Страница 10: ...ocouple failed 6 28 Sensor non volatile memory failed 7 29 Processor board non volatile memory failed 8 30 Main board non volatile memory failed 9 31 Reserved 10 32 Reserved 11 33 Comms board non vola...

Страница 11: ...15 Acknowledge alarm 8 MODBUS address Description 30000 Oxygen concentration value 30002 COe concentration value 30004 Inlet temperature value 30006 Outlet temperature value 30008 Outlet Inlet tempera...

Страница 12: ...rature MODBUS address Description 30100 Calibration progress completion MODBUS address Description 30102 Analog output value mA AOUT1 30104 Analog output value mA AOUT2 30106 Analog output value mA AO...

Страница 13: ...w 40048 Analog output 3 Electrical range high 40050 Analog output 3 Electrical range low 40052 Analog output 3 Engineering range high 40054 Analog output 3 Engineering range low 40056 Analog output 4...

Страница 14: ...couple types 43006 Efficiency calculation 0 Disabled 1 Enabled 43007 Blowback interval 0 Off 1 24 hours 43008 Reserved 43009 Scheduled calibration frequency units 1 Off 1 Daily 2 Weekly 3 Monthly 4301...

Страница 15: ...ht saving 1 First 5 Last 2 Second 3 Third 4 Fourth 43034 Daylight saving end day for daylight saving 1 to 7 where 1 Sunday and 7 Saturday 43035 Daylight saving end month for daylight saving 1 to 12 wh...

Страница 16: ...rrent enable 0 Disabled 1 Enabled 43061 Analog output 4 Source 0 None 6 Efficiency 43062 Analog output 4 Polarity 0 Inverted 1 Non Inverted 43063 Analog Output 4 Failure current enable 0 Disabled 1 En...

Страница 17: ...ag Alarm 1 44096 Alarm 2 tag Alarm 2 44112 Alarm 3 tag Alarm 3 44128 Alarm 4 tag Alarm 4 44144 Alarm 5 tag Alarm 5 44160 Alarm 6 tag Alarm 6 44176 Alarm 7 tag Alarm 7 44192 Alarm 8 tag Alarm 8 MODBUS...

Страница 18: ...18 COM AZ40 MOD EN Rev A Endura AZ40 Oxygen and carbon monoxide equivalent COe analyzer Appendix B Acknowledgements MODBUS is a registered trademark of the Modbus IDA organization...

Страница 19: ...Endura AZ40 Oxygen and carbon monoxide equivalent COe analyzer COM AZ40 MOD EN Rev A 19 Notes...

Страница 20: ...document and in the subject matter and illustrations contained therein Any reproduction disclosure to third parties or utilization of its contents in whole or in parts is forbidden without prior writt...

Отзывы: