Endura AZ series integral and remote transmitter
Combustion oxygen monitor
6 Calibration
IM/AZ20E–EN Rev. F
67
…Calibration Procedures
• New calibration
coefficients
discarded
•
Accuracy
check entry added
to Performance
Log
• Return to
Sensor
Cal.
• New calibration
coefficients saved
Save successful
•
Calibration
entry added to
Performance Log
• Return to
Sensor
Cal.
Save unsuccessful
•
Accuracy
check entry added to
Performance Log
• Display Failure
Calibration completed successfully
1-Pt
Aut
o
Cal
2-Pt
Aut
o
Cal
Man
u
al
Cal
– Of
fset
M
a
n
u
al
Cal –
Fact
or
Manu
al Cal – 2
-Pt
• Return to
Sensor
Cal.
Calibration failed to complete
The reason for failure is displayed:
– stability criteria
not met
– no response to the application
of a test gas detected
– calibration successful but
Save
of
new calibration factor unsuccessful
– newly
calculated calibration factor is outside the limits defined
by
Factor Limit
and / or the newly calculated calibration
offset is outside the
Offset Limit
parameter – see page
.
1-Pt
Aut
o
Cal
2-Pt
Aut
o
Cal
Manu
al
Cal
– Of
fset
Man
u
al
Ca
l –
Fact
or
Manua
l Cal – 2-
Pt
<Cal Type>
O2
1.00
%O2
Cal Offset
0.00 mV
Cal Factor
1.000
Reject
Accept
<Cal Type>
O2 20.95
%O2
Exit
Calibration Failed
Signal Not Stable
Signal Not Stable
No Response
NV Error
Result Out of Bounds