Endura AZ series integral and remote transmitter
Combustion oxygen monitor
5 Programming
IM/AZ20E–EN Rev. F
43
…Calibrate
…Cal. Diagnostics
…Cell Diagnostics
Slow
Recovery
Diag.
Warns if the probe's recovery rate is slower than
expected after a calibration or accuracy check is
measured.
Note
. The recovery rate is measured only if the
difference between the test gas and process gas values
is > 10 % of the oxygen range.
Disabled
Disabled
Enabled
Recovery
Rate
Limit
Sets the limit to be used with the slow recovery
rate diagnostic function.
[0.1 to 10.0 % O
2
/ s]
10.0 % O2 /
s
High
Impedance
Diag.
Warns if the cell's impedance goes above a preset limit
(a high impedance can indicate a faulty or ageing cell).
Disabled
Disabled
Enabled
Performs an impedance check when a
Two point
calibration is selected – see page 39.
Measure
Impedance
With a stable output from the cell, a cell impedance
check can be performed:
Note
. The cell impedance cannot be calculated if the cell
output is less than 20 mV (test gas value greater than
approximately 8 % O
2
@ 700 ºC [1292 °F])
100 K
Parameter
Comment / [Range]
Default
Table 5.7 Calibrate Menu (Continued)