pH / Redox (ORP) transmitter
9 HART Communication
OI/APA592–EN Rev. E
87
12
Read message
Returns the user-defined HART message.
13
Read Tag, Descriptor and
Date
Returns the user-defined HART information:
HART Tag
HART Descriptor
Device installation date
14
Read Primary Variable
Sensor Information
Returns the sensor limits and minimum permissible span of the
sensor.
15
Read Primary Variable
Output Information
PV Alarm selection code:
0 = Low,
1 = High,
PV upper and lower range values.
Damping time.
Write protection status.
Any private label distributor code associated with the device.
16
Read Final Assembly
Number
Returns the user-defined final assembly number.
17
Write Message
Enables a user-defined message of up to 32 characters to be
entered.
18
Write Tag, Descriptor and
Date
Enables the user to assign a user-defined 8-character tag to the
device.
Enables a user-defined 16-character descriptor to be entered.
Enables an installation date to be entered.
19
Write Final Assembly
Number
Enables a user-defined final assembly number to be entered.
This can be used for inventory or maintenance purposes.
Command
No.
Description
Details
Table 9.1 HART Universal Commands (Continued)