Endura ACA592
Toroidal conductivity transmitter
7 Operator Pages and Menus
56
OI/ACA592/TC–EN Rev. B
Temp. Sensor
Temperature sensor parameters.
Temp. unit
Sets the temperature unit (°C or °F) applied to all transmitter temperature parameters and
values.
Temp. Comp. Type
Temperature compensation type.
Manual
Manual
temperature compensation is enabled when the device does not detect any
temperature sensor connection either at power-up or upon execution of
Run TC
Recognition
. When
Temp Comp Type
is
Manual
, the user configurable
Manual Temp
value is used for temperature compensation.
Auto
Enabled when a temperature sensor is detected by the device either at power-up or
when executing the
Run TC Recognition
(see page 56) routine.
Manual Temp
Enabled when
Temp Comp Type
=
Manual.
The value in this parameter is used for temperature compensation during
Manual
temperature compensation mode.
Adjustable range –20 to 150 °C (–4 to 302 °F). Default = 25 °C (77 °F).
Auto TC Option
Enabled when
Temp. Comp Type = Auto
(temperature sensor input is detected).
Options comprise:
Standard KCl
TC Coeff
User Def.
NaOH
NaCl
HCl
H
2
SO
4
TC Coeff
Enabled when a temperature sensor is detected and
Auto TC Option = TC Coeff.
User Def. TC Curve
Enabled when a temperature sensor is detected and
Auto TC Option = User Def. TC.
Enables entry of user-defined temperature compensation curve.
Range = Temp 1 to Temp 6, K1/Kref to K6Kref.
Reference Temp
Sets the required reference temperature.
Temp. Sensor Type
Indicates the type of temperature sensor detected.
Options comprise:
Pt100
Pt1000
Run TC Recognition
Starts the auto recognition process for the sensor.
TC Recognition Stat
Indicates if the temperature sensor was recognized / not recognized, either during startup or
after execution of the
Run TC Recognition
function.
Device Setup Level
Parameter
Description