![ABB Endura ACA592-TC Скачать руководство пользователя страница 57](http://html1.mh-extra.com/html/abb/endura-aca592-tc/endura-aca592-tc_operating-instructions-manual_3052232057.webp)
Endura ACA592
Toroidal conductivity transmitter
7 Operator Pages and Menus
OI/ACA592/TC–EN Rev. B
55
7.3.3 Device Setup
Used to set the sensor / sensor input, run temperature sensor
/ sensor recognition and obtain transmitter information.
Note
. Required level of access =
Advanced
.
Device Setup Level
Parameter
Description
Main Sensor
Analyzer Type
Sets the analyzer to a measurement type:
Conductivity
Concentration
Conc. Soln.
Enabled if
Analyzer Type = Concentration
.
Sets the concentration solution:
0–18% HCI, 0–20% H
2
SO
4
, User Def., 0–1% NaOH or 0–20% NaC1.
Conc. Unit.
Enabled if
Analyzer Type = Concentration
.
Selects the concentration unit:
%, ppm, µg/l, ppb, mg/l or none
Conc. Curve Name
Enabled if
Analyzer Type = Concentration
.
A user-defined name.
Press the
key to enter the required characters for the name as detailed on page 40.
Conc. Curve Table
Enabled if
Analyzer Type = Concentration
.
Enables entry of 6-point conductivity / concentration curve for user-defined concentration
calculation.
Range = X1 to X6 conductivity in uS/cm, Y1 to Y6 concentration
3
Menu
Exit
Select
Device Setup