Error/Action
Start-up Screen
Displayed at system start-up.
Calibration error
Turn mains power off and on again
(if the error persists contact the
Service Organization).
Configuration error
The configuration and/or setup data
for the instrument is corrupted. Turn
mains power off and on again (if
e r r o r p e r s i s t s , c h e c k
configuration/setup settings).
A to D Converter fault
The analog to digital converter is
not communicating correctly.
Sensor input out of range
Option board error
Communications to the option
board have failed.
Display
To Clear Display
Clears automatically (after
3s) to display software
version/issue – see Fig.
1.1.
Press the key
Press the key
Turn mains power off and
on again. If the problem
persists, contact the
Service Organization
Restore valid sensor input.
Check sensor output.
Contact the Service
Organization
IP.FAIL
CAL.Err
CFG.Err
A.d. Err
OPt.Err
tESt
…1
DISPLAYS AND FUNCTION KEYS
1.4
Error Messages