Preparing the EtherNet Cat5e cable
1
90 mm (3.54 in)
2
39 mm (1.54 in)
3
60 mm (2.36 in)
4
Tin 10 mm of the end of the
braided shield of the cable
Figure 59: Preparing the EtherNet Cat5e cable
Ground the Ethernet connection cable
Connect the outer shield of the Ethernet cable to the screw
terminal.
A
Screw terminal
Figure 60: Ground the Ethernet connection cable
M12 plug (option)
Various options for M12 plugs are available from the model code:
• Flowmeter equipped with 1 × M12
(with 4 wires, connecting to Port 1)
• Flowmeter equipped with 2 × M12
(with 4 wires each, connecting to Port 1 & 2)
• Flowmeter equipped with 1 × M12
(with 8 wires, connecting to Port 1 & 2)
These options allow for connection to different network
topologies:
Topology 4
wires
4 wires
4 wires
8 wires
1 x M12
(4 wires)
2 x M12
(4 wires)
1 x M12
(8 wires)
Star Y
Y
Y
Ring or daisy- chain
N
Y
N
PoE N
N
Y
Electrical connections
For the internal wiring inside the transmitter and the related
pinout within the M12 Plug, refer to table below:
Wiring inside the transmitter
M12 Plug pin
Color Ethernet card
port/pin
M12 plug with 4 wires
1
Yellow
Port 1 X1
2
Orange
Port 1 X2
3
White
Port 1 X3
4
Blue
Port 1 X4
M12 plug with 8 wires
1
White
Port 1 X1
2
Blue
Port 1 X2
3
Brown
Port 1 X3
4
Green
Port 1 X4
5
Pink
Port 1 X5
6
Yellow
Port 1 X6
7
Grey
Port 1 X7
8
Red
Port 1 X8