INSTALLATION
2.2.2
Fitting the Adaptor Board
(Fig. 2.3 and 2.4)
a) Turn off the power to the MagMaster.
b) Record the position of the sensor cable
wiring and remove the wiring from the
terminals. Slacken the terminal screws
by at least 6 turns.
c)
Carefully ease the sensor wiring to one
side and arrange the adaptor board so
that the extended pins fit into the
terminals previously occupied by the
sensor wiring – see Fig.2.4.
d) Fit and tighten the securing screw to the
corner of the adaptor board.
e) Tighten the associated terminal block
screws.
Fig 2.3 Normal Sensor Wiring in the Transmitter
Fig 2.4 Fitting the Adaptor Board
DS 2
IC 2
CD 1
CD 2
SIG GND
DS 1
SIG 1
Grey Coax
Pink Inner (SIG 2)
Screen (DS 2)
Red
(CD 1)
Yellow
(CD 2)
White Coax
Blue Inner (SIG 1)
Screen (DS 1)
Ground Wire
(Safety Earth)
Violet
Adaptor
Board
ALARM 1
F OUT
A
ALARM 2
F OUT
B
PLS OV
PLS OV
EXT
I/P+
EXT
I/P-
TX- SIG
RX- SIG
TX+ SIG
RX+ SIG
OVC
N
+ L