Aztec ATS430 | Turbidity sensor
|
OI/ATS430–EN Rev. H 21
6.4.6 Collection complete
This is the second step of the calibration. Once the suspended
solids content of the sample has been measured in the
laboratory, the value can be entered into the transmitter. Note
that this sample must correspond to the last sample collection
step performed, otherwise the calibration may not be correct.
1.
To start the
Collection Complete
procedure from the
Calibration Type
screen, use the
/
keys to select
Collection Complete
and press the
key (below the
Edit
prompt).
The
Suspended Solids / New Process
screen is displayed:
2. Press the
key (below the Edit prompt).
The
New Process
screen is displayed:
3. To start collection for a new process (when the sensor is
installed for the first time in a new process or when the
calibration needs to be reset) use the
/
keys to
select
Yes
.
To retain the memory of the previous calibrations (adaptive
calibration to fine tune the existing suspended solids
calibration) select
No
.
4. Press the
key (below the
OK
prompt).
5. Press
key (below the
Next
prompt).
The
Collection Complete
screen is displayed:
This screen displays the turbidity recorded when the
sample was taken (read-only PV field) and a field (TSS) to
enter the suspended solids value measured in the lab.
6. Use the
/
keys to enter the suspended solids value
(ensure the value entered is in the same units as those
displayed in the screen) and press the
key (below the
Continue
prompt) when complete.
A new calibration coefficient based on the value entered,
and the previous values if using adaptive calibration, is
calculated.
The calibration is now complete.
7.
Press the
key (below the
Exit
prompt) to return to the
main menu.
Suspended Solids
Next
Edit
Calibration Type
Collection Complete
Suspended Solids Ca
Next
Edit
New Process
No
New Process
Cancel
OK
Yes
No
Collection Complete
Next
Continue
PV
2000 NTU
TSS
000000 mg/l