background image

AquaMaster 3 
Electromagnetic flowmeter 

 

MODBUS Tables Supplement COI/FET2XX/MOD/TBL-EN 

 24 

 

MODBUS Protocol 

Tables of Settings

 

0

 

Special

 

1

 

N/A

 

2

 

PFA

 

3

 

PTFE

 

4

 

FEP

 

5

 

ABR

 

6

 

Rubber (WRC)

 

7

 

Polyurethane

 

8

 

HT Rubber

 

9

 

Polypropylene

 

10

 

PES

 

11

 

P E

 

12

 

PPS

 

13

 

Viton

 

14

 

St St

 

15

 

C St

 

16

 

Brass

 

17

 

Gold

 

18

 

Titanium

 

19

 

Hast B

 

20

 

Hast C

 

21

 

Plat

 

22

 

Tant

 

23

 

?

 

Table A.1 Materials

 

0

 

N/A

 

1

 

Class1

 

2

 

Class2

 

3

 

MID Class1

 

4

 

MID Class2

 

Table A.2 Calibration Types 

Содержание AquaMaster 3

Страница 1: ...ation information and MODBUS Protocol Register Allocation tables and is applicable to product firmware version WAJC2103 v01 00 00 or above Detailed specifications and recommendations for using and imp...

Страница 2: ...committed to teamwork high quality manufacturing advanced technology and unrivalled service and support The quality accuracy and performance of the Company s products result from over 100 years experi...

Страница 3: ...ister Usage 4 1 3 1 RW 1 Bit 4 1 3 2 RO 1 Bit 5 1 3 3 RO 8 16 Bit 5 1 3 4 RO String 6 1 3 5 RO 32 Bit 7 1 3 6 RO Float 8 1 3 7 RW 8 16 Bit 8 1 3 8 RW String 10 1 3 9 RW 32 Bit 12 1 3 10 RW Float 13 Ap...

Страница 4: ...dered when implementing a MODBUS master to communicate with AquaMaster 3 Transmission mode Remote Terminal Unit RTU slave Response time out 1s Turnaround delay 800ms Interface power up considerations...

Страница 5: ...for each Register Address Range and Appendix A provides the Register allocation tables Register Address Range Applicable Function Codes AquaMaster Data Type Description 0001 0500 01 Read Coils 05 Writ...

Страница 6: ...le bit types Read Coils 1 to 14 Read all RW 1 Bit items Request Message Hex Response Message Hex Error Message Hex Address 01 Address 01 Function 81 Function 01 Function 01 Exception Code Exception Co...

Страница 7: ...Variable Meaning 0 501 0 007 Flow sensor is not reverse wired 1 502 1 008 Flow sensor is a Probe 2 503 1 416 MID read only switch is Active 3 to 7 Not Used 1 3 3 RO 8 16 Bit Read only access to 8 and...

Страница 8: ...Each AquaMaster 3 String type requires 18 MODBUS Registers 36 bytes Consequently RO String Register addresses are assigned at intervals of 18 and the Request Message must provide Quantity of Inputs i...

Страница 9: ...ters 4 bytes Consequently RO 32 Bit Register addresses are assigned at intervals of 2 and the Request Message must provide Quantity of Inputs in multiples of 2 32 Bit quantities are arranged in big en...

Страница 10: ...g Address Hi 13 Byte count 04 Starting Address Lo a8 H Register Value Hi 41 Quantity of Inputs Hi 00 H Register Value Lo 20 Quantity of Inputs Lo 02 L Register Value Hi 00 CRC f4 L Register Value Lo 0...

Страница 11: ...Value Hi 00 Register Value Hi 00 Register Value Lo 0e 14 Register Value Lo 0e 14 CRC ae CRC ae CRC 0b CRC 0b Write Multiple Registers 8019 var125 Trip point for high flow 15 Request Message Hex Respo...

Страница 12: ...nse Message Hex char Error Message Hex Address 01 Address 01 Function 83 Function 03 Function 03 Exception Code Exception Code Starting Address Hi 27 Byte count 24 Starting Address Lo e8 byte0 41 A Qu...

Страница 13: ...Address Lo d6 Starting Address Lo d6 Quantity of Registers Hi 00 Quantity of Registers Hi 00 Quantity of Registers Lo 12 Quantity of Registers Lo 12 Byte count 24 CRC ab byte0 46 F CRC 48 byte1 6c l...

Страница 14: ...antity of Registers in multiples of 2 32 Bit quantities are arranged in big endian order Read Holding Registers 12003 to 12004 var224 Totaliser Forward Request Message Hex Response Message Hex Error M...

Страница 15: ...s Hi 00 Register Value H Lo 66 Quantity of Inputs Lo 02 Register Value L Hi 66 CRC 0a Register Value L Lo 67 0 9000000 CRC 8c CRC 7c CRC 72 Each AquaMaster 3 RW Float type requires 2 MODBUS Registers...

Страница 16: ...al on AquaMaster 3 LCD display 2 053 DisplayReverseTotal 0 Off 1 On 3 054 DisplayNetTotal 0 Off 1 On 4 055 DisplayTariffA 0 Off 1 On 5 056 DisplayTariffB 0 Off 1 On 6 059 DisplayFlowRate 0 Off 1 On 7...

Страница 17: ...Sensor settling time mains power 1008 034 ABB Reserved 2 1009 035 ABB Reserved 2 1010 036 ABB Reserved 1 1011 106 CalibrationType 1 0 to 4 See Table A 2 Calibration Types 1012 148 ABB Reserved 1 1013...

Страница 18: ...2109 199 TransmitterType 25 AquaMaster 3 2127 200 Manufacturer 25 2145 201 Address1 25 2163 202 Address2 25 2181 203 Address3 25 2199 204 Address4 25 2217 210 CalibrationDate 11 Transmitter calibrati...

Страница 19: ...03 002 FlowSensorSoftwareRev 4 4005 003 FlowSensorHardwareRev 4 4007 172 PressureSoftwareRev 4 4009 173 PressureHardwareRev 4 4011 205 HardwareRevisionLevel 4 4013 207 TransmitterUniqueID 4 Transmitte...

Страница 20: ...tedBore 4 mm 5035 238 ISpan 4 Factory coil current span factor 5037 239 IZero 4 Factory coil current zero correction 5039 243 CoilCurrent 4 A 5041 258 PulseOutput 4 Hz 5043 294 SignalCalHighCurrent 4...

Страница 21: ...to 19 See Note 1 8016 117 FlowCutOffLow 1 0 to 255 mms 1 y Factory flow cut off battery mode 8017 118 UserFlowCutOff 1 0 to 255 y User s flow cut off setting 8018 119 PressureUnits 1 0 to 7 See Note...

Страница 22: ...15 V 8054 326 ABB Reserved 1 8055 327 ABB Reserved 1 y 8056 334 SoftwareResetCount 2 8057 335 WatchdogResetCount 2 8058 336 SensorCalTxVersion 1 y 8059 339 ABB Reserved 2 8060 342 LastTxSoftwareVer 1...

Страница 23: ...g 25 10271 186 Pressure1stFactory 11 See Table A 4 Date Format 10289 187 PressureLastFactory 11 See Table A 4 Date Format 10307 188 PressureCertNo 25 10325 189 PressureContractNo 25 10343 196 Pressure...

Страница 24: ...itted unless MID active 12005 225 TotaliserRev 4 Totaliser units y Write value of 0 permitted unless MID active 12007 226 TotaliserNet 4 Totaliser units y Write value of 0 permitted unless MID active...

Страница 25: ...15 110 CustomVelocityUnits 4 per ms 1 13017 113 CustomFlowUnits 4 per m3 s 1 13019 115 FlowUpperRange 4 Flow units y 13021 116 FlowLowerRange 4 Flow units y 13023 120 CustomPressureUnits 4 per mmHg 13...

Страница 26: ...ttings 0 Special 1 N A 2 PFA 3 PTFE 4 FEP 5 ABR 6 Rubber WRC 7 Polyurethane 8 HT Rubber 9 Polypropylene 10 PES 11 P E 12 PPS 13 Viton 14 St St 15 C St 16 Brass 17 Gold 18 Titanium 19 Hast B 20 Hast C...

Страница 27: ...OW 2 Battery Power FAIL 3 Mains Power ON 4 Mains Power OFF 5 Renewable Power OK 6 Renewable Power LOW 7 Renewable Power OFF Table A 3 External Power Status Modbus register 8030 AquaMaster 3 parameter...

Страница 28: ...AquaMaster 3 Electromagnetic flowmeter MODBUS Tables Supplement COI FET2XX MOD TBL EN 26 MODBUS Protocol Notes...

Страница 29: ...AquaMaster 3 Electromagnetic flowmeter MODBUS Tables Supplement COI FET2XX MOD TBL EN 27 MODBUS Protocol...

Страница 30: ...AquaMaster 3 Electromagnetic flowmeter MODBUS Tables Supplement COI FET2XX MOD TBL EN 28 MODBUS Protocol...

Страница 31: ...nalytics Instrumentation pH Conductivity and Dissolved Oxygen Transmitters and Sensors Ammonia Nitrate Phosphate Silica Sodium Chloride Fluoride Dissolved Oxygen and Hydrazine Analyzers Zirconia Oxyge...

Страница 32: ...t without prior notice With regard to purchase orders the agreed particulars shall prevail ABB does not accept any responsibility whatsoever for potential errors or possible lack of information in thi...

Отзывы: