26
Installation Instructions ACF-NT, ACF-NT V0309
42/23-571 EN Rev. 5
Connection Plan: Power Supply
(3
)
(1
)
1
-X10
4
2
3
PE
L2
L3
L1
PE
N
230/400V
+/-10,
48...62Hz
-X20
2
1
PE
PE
N
L1
l=
1,5m:
600W
Pt
100
-X01
11
10
12
l=
2m:
800W
-X02
2
1
3
l=
1m:
400W
Pt
100
250W
-X01
13
15
14
-X02
4
6
5
7
Pt
100
-X01
19
16
22
-X02
18
17
90W/m
-X120
8
9
N
L
PE
~
~
(120/208V
+/-10,
48...62Hz)
N
-E02
L
PE
Sensor
Temperature
Cable Gland M20
Analyzer
Cabinet
ACF-NT
UPS Power Supply
Cable Gland M32
Cable 5x6 mm2 or 5xAWG8
Fuse-protection external 3x20A (3x25A)
(Supplied by Customer)
(Cable Dia. Range 7-15mm)
Power Supply
Cable 3x2,5 mm2 or 3xAWG14
Heater Sample Tube Type 42(-E11)
Cable Gland M25
(Cable Dia. Range 6-13mm)
Fuse-protection external 20A
Cable 3x 2,5 mm2 0r 3xAWG14
(Supplied by Customer)
(Cable Dia. Range 6-13mm)
Cable Gland M25
Sample
Cable 3x0,75 mm2 or 3xAWG16
Cable Gland M20
(Cable Dia. Range 5-10mm)
Connection via Terminal Box on-site
Heater Filter Unit (-E12)
Cable 3x1,5 mm2 or 3xAWG16
Cable Gland M20
(Cable Dia. Range 5-10mm)
Sample
Power
Supply
Heating
Temperature
Sensor
Heating
40m
-
3phase
Fuse-protection
15m
-
1phase
Fuse-protection
(Standard)
CAUTION
:max.
Length
of
Sample
Gas
Line
at
120/208V
:
(Cable Dia. Range 5-10mm)
(Cable Dia. Range 5-10mm)
Connection via Terminal Box on-site
Cable Gland M20
(Cable Dia. Range 6-13mm)
Cable Gland M25
Cable 3x2,5 mm2 or 5x2,5mm2
Heated Sample Line TBL01 (-E13)
Air Purifier (-E05)
Plug Power Supply (-G05)
via delivered
Temperature
Sensor
Heating
in
Side
W
all
Socket
T
OO
HIGH
/
6mA
LEAKAGE
CURRENT
W
ARNING
!
in
Cabinet
Light
Socket
230V AC or 120V AC, 48...62Hz, max.5A
CAUTION
:max.
Length
of
Sample
Gas
Line
at
230/400V
:
35m
-
1phase
Fuse-protection
(Standard)
60m
-
3phase
Fuse-protection
Cable 3x0,75 mm2 or 3xAWG16
Cable 3x0,75 mm2 or 3xAWG16
Cable 3xAWG14 or 5xAWG14