ABB 4234 600 Скачать руководство пользователя страница 7

5

To gain access, remove the four
screws and lift off the cover.

Fig. 4.2 PSU Cover Removal

TB1

T1

F2 F3

6

8

10

230 V

115 V

L    N    E

115 V a.c.

or

230 V a.c.

Supply

See Caution opposite

Tighten Cable Gland

M5 Earth Stud

Fig. 4.4 Connecting Outputs

Fig. 4.3 Connecting to a.c. Supply

4 ELECTRICAL INSTALLATION

To Katharometer
(see manual for
termination details)

Tighten Cable Gland

TB2

+

F1

B

A

C

X

D

E

d.c. Output
Terminal Block

Cable Clamp

Output Current (mA) Links

500

A to X

400

B to X

350

C to X

250

D to X

180

E to X

Current Output Link

To Katharometer
temperature control
circuit
(see manual for
termination details)

AUX 10V

+

4.2

Accessing the PSU Terminals

Warning. If the PSU is cubicle mounted or has its

cover removed for test purposes, a.c. mains supply
voltage is exposed at certain locations within the
assembly when supplies are connected. Thus, there is a
risk of electrocution.

All normal precautions must be taken to avoid the risk of

accidental electrical shock during installation and
maintenance of the equipment.  Ensure that the supply
cable is disconnected at its source before touching any
electrical connections.

Caution. Ensure that the correct version of the unit

has been supplied for the supply voltage available.  A 115 V
unit cannot be used with a 230 V supply or vice versa.

Check that the current output meets the katharometer

input requirements (see katharometer data label) – see
Fig. 4.4. If in doubt contact ABB Limited.

Remove the power supply unit cover to obtain access to the
separate labelled terminal blocks – see Fig. 4.2.

4.3

Electrical Connections

The mains terminal block, TB1, is situated next to the transformer,
T1, and fuses, F2 and F3.

Connect the a.c. mains input cable as in Fig. 4.3.

 Note. Earth bonding to be made via M5 Earth Stud.

Connect the d.c. output and 10 V cables as in Fig. 4.4.

Содержание 4234 600

Страница 1: ...4234 600 Series Power Supply Unit Instruction Manual Models 4234 600 4234 601 ...

Страница 2: ...rational conditions result in degraded process system performance leading to personal injury or death Therefore comply fully with all Warning and Caution notices Information in this manual is intended only to assist our customers in the efficient operation of our equipment Use of this manual for any other purpose is specifically prohibited and its contents are not to be reproduced in full or part ...

Страница 3: ...rviceable parts Two identical fuses L N are mounted within the unit on the mains input A third fuse is connected in the d c line circuit of the transformer A fourth fuse is connected in the auxiliary 10 V d c line The L N fuses are a high breaking capacity type and must only be replaced with identical types See Section 2 for rating 1 INTRODUCTION Fig 1 1 Location of Internal Component TB1 T1 L N E...

Страница 4: ...terconnecting cable 2 Ω max Ambient Temperature Range 20 to 55 C 2 SPECIFICATION Permissible Supply Voltage Variation 15 Frequency Range 46 to 64 Hz Output Regulation Within 0 8 for a Load variation 15 b Supply variation 6 c Ambient temperature 10 C Ripple Less than 17 5 mV across 10 Ω load peak to peak Stability Within 0 7 of initial setting over a period of one month with load resistance supply ...

Страница 5: ...ore commencing installation check that the unit supplied is suitable for the supply voltage available see Section 2 and the data plate on the unit cover Fig 3 2 140 CTRS 110 CTRS 170 111 120 All holes 7 0 diameter 160 83 20 20 30 30 30 60 DIM Ensure a clearance of at least 100 mm at both ends of the unit for acces to the cable glands and to minimise cable bends Mains Cable Entry Output to Temperat...

Страница 6: ...nections for typical gas analysis systems 4 1 1 Katharometer Connections Cable connecting the constant current d c supply from the PSU to the katharometer should be selected so that the maximum cable resistance is less than 2 ohms This limits the maximum cable length between the katharometer and its associated PSU Note that the cable gland will accept a cable between 5 to 9 mm in diameter 4 1 2 Te...

Страница 7: ... assembly when supplies are connected Thus there is a risk of electrocution All normal precautions must be taken to avoid the risk of accidental electrical shock during installation and maintenance of the equipment Ensure that the supply cable is disconnected at its source before touching any electrical connections Caution Ensure that the correct version of the unit has been supplied for the suppl...

Страница 8: ...Centre or returned to the factory for repair Fig 5 1 Changing Fuses TB1 T1 L N E F2 F3 6 8 10 230 V 115 V TB2 F1 B A C X D E F4 AUX 10V Switch off and disconnect the mains supply 1 2 Remove the four screws and lift off the cover Replace fuse s 3 Note Suitable fuses may be obtained from ABB with the following part numbers F1 0231555 400 mA cartridge F2 F3 0231577 250 mA 1500 A hbc cartridge F4 0231...

Страница 9: ...7 NOTES ...

Страница 10: ...8 NOTES ...

Страница 11: ...al Analytics Instrumentation pH conductivity and dissolved oxygen transmitters and sensors ammonia nitrate phosphate silica sodium chloride fluoride dissolved oxygen and hydrazine analyzers Zirconia oxygen analyzers katharometers hydrogen purity and purge gas monitors thermal conductivity Customer Support ABB Ltd provides a comprehensive after sales service via a Worldwide Service Organization Con...

Страница 12: ...ained herein without notice Printed in UK 01 02 ABB 2002 ABB Ltd Oldends Lane Stonehouse Gloucestershire GL10 3TA UK Tel 44 0 1453 826 661 Fax 44 0 1453 827 856 ABB Inc 2175 Lockheed Way Carson City NV 89706 USA Tel 1 0 775 883 4366 Fax 1 0 775 883 4373 ABB has Sales Customer Support expertise in over 100 countries worldwide www abb com ...

Отзывы: