42
ELECTRICAL CAL
Calibrate No
xxx.x
mV Zero (0.00mV)
xxx.x
mV Span
Adjust RTX Zero
Adjust RTX Span
Yes
No
Yes
10.3.4
Electrical Calibration Page (LOWER DISPLAY) – AK101 only
Notes
.
1) The 4689 instruments incorporate a two-point calibration sequence requiring both
zero and span inputs for a calibration. It is not possible to adjust the range zero or the
range span scale points independently.
2) The instrument is fully calibrated before despatch and should not normally require
further calibration.
Select Calibration
Select the calibration requirement using the
or
keys.
Calibrate No
(default) skips to Adjust RTX Zero frame.
Calibrate Yes
enables zero and span electrical calibrations to be carried out.
Advance to next parameter
Calibration Range Zero (0% Air in CO
2
)
Proceed as described in Section 8.3 Calibration, but apply a signal input equivalent to
%AIR IN CO2 range zero (0.00mV).
Allow the instrument to stabilize.
.
Advance to next parameter
Calibration Range Span (100% Air in CO
2
)
Apply a signal input equivalent to:
%AIR IN CO
2
range span (10.00mV)
N
2
range span (1.00mV)
Argon range span (10.00mV)
Allow the instrument display to stabilize.
Advance to next parameter
Adjust Retransmission Zero
Adjust the retransmission zero (e.g. 4.00mA) to the appropriate zero value.
The retransmission zero signal is either 85% or 80% H
2
in Air as selected in
Set Up
Outputs Page
.
Allow the output signal to stabilize.
Advance to next parameter
Adjust Retransmission Span
Adjust the retransmission span (e.g. 20.00mA) to the appropriate maximum value.
The retransmission span signal corresponds to 100% H
2
in Air.
Allow the output signal to stabilize.
Return to
Operating Page
.
...10