< 45 cm not preferred

Up to 45 cm 

below is ok

Ideal

•  Open the accessory box and use this list to make sure you received all  

  components of the Vet-Scan HM5 system.
•  VetScan HM5 Analyzer.
•  VetScan HM5 Operator’s Manual.
•  External power supply.
• Mini-keyboard
•  Four sample tube adaptors.
•  Reagent Tubing Kit.

Selecting a Location 

The analyzer must be installed in a suitable location.

 

•  The location should be flat, level, sturdy, vibration-free, and as dust-free as possible.
•  Leave at least 50 cm of clearance on all sides of the instrument to allow for  

  adequate airflow and access to connectors.
•  Maintain a minimum of 20 cm behind the analyzer’s rear panel for heat dissipation  

  and tube clearance.
•  Leave enough space for the reagent pack.

Connecting the Reagent Pack

•  Place the reagent pack near the analyzer, either beside  

  or behind the instrument.
•  For best results, place the reagent pack at the same level as  

  the analyzer. If you have to place the reagent pack on a lower  

  level, make sure that it is no more than 45 cm below the  

  level of the analyzer’s reagent inlets, as measured from the  

  bottom of the reagent pack to the bottom of the analyzer.

Note:

 

Do not place the reagent pack on a higher level.

 

Closer to Care.

Содержание VetScan HM5

Страница 1: ...QUICK REFERENCE GUIDE Easy operation fast results Closer to Care...

Страница 2: ...re the shipping box accessory box and packaging materials in case you need to use them later Place the analyzer on a level surface Use the latch to open the side door on the unit as shown Inspect the...

Страница 3: ...sides of the instrument to allow for adequate airflow and access to connectors Maintain a minimum of 20 cm behind the analyzer s rear panel for heat dissipation and tube clearance Leave enough space...

Страница 4: ...ould snap into each other when properly connected Connect waste bottle Initialization of the HM5 Connect Power Supply Install paper into the built in printer Turn on the analyzer by using the power sw...

Страница 5: ...age reagent status Exit select shutdown options and manage user accounts General Settings select analyzer options and language Date and Time change date and time options Measurement select units set n...

Страница 6: ...d whole blood sample into the appropriate tube adapter on the sample rotor and remove the cap Touch the Run button If your HM5 is set for autoprint your results will be printed automatically If not pr...

Страница 7: ...grams If the analyzer is not set to print results automatically press Print to print the results Results falling outside of the specified species reference range are marked as follows Results above th...

Страница 8: ...e the analyzer has been powered on and off again If the analyzer is in standby mode touch the measure icon Touch the Run button If a blank is necessary the Blank needed screen appears Touch OK to conf...

Страница 9: ...te fields b Enter by USB pen drive the QC values file can be loaded by inserting the USB drive into a USB port and touching Load c Enter by Barcode reader press Load Datamatrix on the Quality Control...

Страница 10: ...Changing the Reagent Pack In the Maintenance menu select Reagent status Select Replace and then Change pack Select Prime All to prime the reagent tubing...

Страница 11: ...assay values from the calibrator package insert and press Accept Mix the Normal Control well by inverting it ten times uncap the vial and place it into the vial adaptor Touch the Run button When the...

Страница 12: ...h a damp lintfree cloth With each change of reagent pack Perform Vetscan HM5 bleach cleaning procedure Enzymatic Cleaning An enzymatic cleaning is recommended after approximately every 100 tests or at...

Страница 13: ...ut less than 10 days follow this procedure 1 Touch the Exit icon 2 Select shut down then press accept 3 Power off the analyzer when it emits a tone More than 10 days Follow the instructions below if t...

Страница 14: ...se resistant RBCs in canine samples Re run sample at higher lyse volume X EOX blank high Check that Lyse2 is room tempera ture Y EOS clogging Run 3 automated self cleanings and repeat analysis Z Noise...

Страница 15: ...Notes Closer to Care...

Страница 16: ...ENG Rev A Closer to Care ABAXIS Europe GmbH Bunsenstr 9 11 64347 Griesheim Germany E mail abaxis abaxis de www abaxis de ABAXIS Inc 3240 Whipple Road Union City CA 94587 USA Phone 1 510 675 65 00 www...

Отзывы: