Electromagnetic immunity
Guidance and manufacturer’s declaration – electromagnetic immunity
The Platinum Series is intended for use in the electromagnetic environment specified
below. The customer or the user of the Platinum Series system should assure that it is
used in such an environment.
Immunity test
IEC 60601
test level
Compliance level Electromagnetic environment -
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV
contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30 %.
Power frequency
(50/60 Hz)
Magnetic Field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
Radiated RF IEC
61000-4-3
3 V/m
80 MHz to
2.5 GHz
3 V/m
Portable and mobile RF communications
equipment should be used no closer to
any part of the Platinum Series system,
including cables, than the recommended