-27-
calibration coefficient
К
= 1.
To recalibrate Qulon-12m you need: a) determine the
value of calibration coefficient for your batteries (see section
7) and b) adjust Qulon for work with this coefficient (save this
coefficient in Qulon’s memory).
To change calibration coefficient enter main menu,
select item “options”, enter options menu and select item
“Calibrate”. After entering this item, you will see the current
coefficient. You will be asked if you want to change coeffi-
cient.
After “no” answer (lower button), you will return to the
main menu.
After “yes” answer (upper button), you will see the
screen with current coefficient. Pressing upper button in-
creases coefficient (you see the current value on the screen).
Pressing lower button decreases coefficient (again you see
on the screen the new value). Holding button causes fast
coefficient changing. After the desired coefficient is on the
screen, disconnect Qulon from the battery. The coefficient is
saved in memory. It will be used during further capacity esti-
mations until it is changed once more. Any time you test a
battery, you see the coefficient value in the right top corner of
the screen.
6.4. Filtration
Qulon-12m is equipped with a very effective filter for
noise suppression. Nevertheless, under some circumstances
(very low signal from the battery, very