8

6. Calibration of the balance 
 

Display shows

 

 

CAL 

Press the „

change“ 

button, after a few seconds the value of 

the calibration weight appears in the display. 

 

Place the calibration weight on the middle of the scale pan. 

 

Display disappears. 

 

After a few seconds appears --- 200,000 g 

 

 
 

7. Getting correct measuring values 

 
7.1.  Very  important  is  a  good  preparation  of  the  sample  material.  It  should  be 
homogeneous  and  no  bigger  parts  should  be  there.  Parts  of  more  than  5  mm 
diameter can stay wet in the middle and incorrect values are the results. 
7.2. Different material should be crushed or rasped and liquids should be well stirred. 
7.3. Use a cold drying pan for every new measurement. A cold sample on a hot pan 
will cause a weight loss even before the start of the measurement. 
7.4. We recommend a sample weight of at least 4 g. The greater the sample weight, 
the better reproducible your measurement will be. 
7.5  Try  to  find  out  the  best  drying  temperature  to  reach  short  measuring  times  and 
good results. 
7.6 It is recommended to have some reference values by using the standard method 
with drying cabinet and separate balance. 
7.7.  For  drying  liquids  or  sticky  materials  we  recommend  to  use  aluminium  foils  as 
sample  trays  to  keep  the  stainless  steel pans  free  from  sample  remains.  See list of 
accessories. 
 
 

8. Cleaning and maintenance 

 
8.1  Please  transport  the  instruments  very  careful.  The  built-in  balance  is  very 
sensitive against shocks. 
8.2 Clean the heater and the surfaces of the instrument only with a wet duster. Plug 
out the mains cable before cleaning the radiator. 
8.3 Any maintenance should only be done by authorized persons. 
 

Содержание 2001 Series

Страница 1: ...Instruction Manual Series 2001 instruments p a a p instruments e K Kluckhofer Weg 61 D 32657 Lemgo Tel 49 05232 9778 0 Fax 49 05232 9778 20 Email ap instruments t online de Internet www apinstruments...

Страница 2: ...n a vibration free surface Parts of sample material shall not fall between the pan holder and the housing that will hinder the balance Never touch the hot radiator there is a risk of injury Pull out t...

Страница 3: ...20 sec interval 1 min start interval That means the instrument will start the drying process wait 1 minute after the start start interval until the weight loss is registered by the processor and it wi...

Страница 4: ...e instruments into the automatic mode 0 000 g press SET 2 seconds CON 12 3 or press change 12 3 blinks set time with button change confirm with SET 10 3 blinks 10 0 blinks press SET T I M E R is s e t...

Страница 5: ...e Settings level DISPLAY shows SETTINGS can be changed by using the Change button MEANING PNr 0000 1000 9000 Standard is no sample no Sample no 0001 0002 IA 0 6 12 30 Sek 1 2 4 min Printing in interva...

Страница 6: ...accuracy of measurement of approx 0 1 this parameter has to be set correctly E g a sample weight of approx 4 8 g needs a setting to 4 mg because 4 mg is 0 1 of 4 g Sample weight Recommended setting 4...

Страница 7: ...drying temperature can be set by adjusting the menu item Heat to an exact value of energy This means of 230 Volt power Of course there is a certain connection between the power and the temperature Ra...

Страница 8: ...weight loss even before the start of the measurement 7 4 We recommend a sample weight of at least 4 g The greater the sample weight the better reproducible your measurement will be 7 5 Try to find ou...

Страница 9: ...contact with the housing The instrument shows other irregular reactions e g the radiator cannot be regulated or the parameters are lost Switch off and on the instrument Check if the correct type of i...

Страница 10: ...xistent The data correspond to the V 24 agreements for serial asynchronous data transmission The structure of a data word follows from the following sketch Start Start LSB MSB Stop 0V 12V Bit 0 1 6 Bi...

Отзывы: