23
Table 4
:
Guidance and manufacturer’s declaration - electromagnetic Immunity
The [Code SI] is intended for use in the electromagnetic environment specified below.
The customer or the user of the Pulse Oximeter should assure that it is used in such an environment
Radiated RF
IEC61000-4-3
(Test specifications
for ENCLOSURE PORT
IMMUNITY to
RF wireless communi-
cations equipment)
Test
Frequency
(MHz)
Band a)
(MHz)
Service a)
Modulation b) Modulation b)
(W)
Distance
(m)
IMMUNITY
TEST LEVEL
(V/m)
385
380–390
TETRA 400
Pulse
modulation b)
18 Hz
1,8
0,3
27
450
430–470
GMRS 460,
FRS 460
FM c)
± 5 kHz
deviation
1 kHz sine
2
0,3
28
710
704–787
LTE Band 13,17
Pulse
modulation b)
217 Hz
0,2
0,3
9
745
780
810
800–960
GSM 800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
Pulse
modulation b)
18 Hz
2
0,3
28
870
930
1720
1700–1990
GSM 1800;
CDMA 1900;
GSM 1900;
DECT;
LTE Band 1, 3,4,25;
UMTS
Pulse
modulation b)
217 Hz
2
0,3
28
1845
1970
2450
2400–2570
Bluetooth,
WLAN,
802.11 b/g/n,
RFID 2450,
LTE Band 7
Pulse
modulation b)
217 Hz
2
0,3
28
5240
5100–5800
WLAN 802.11
a/n
Pulse
modulation b)
217 Hz
0,2
0,3
9
5500
5785