9.  Troubleshooting 

Problem Possible 

Reason 

Recommended 

Action 

Batteries are empty. Replace all batteries with new ones. 

Nothing 
appears in the 
display, even 
when the power 
is turned on. 

Battery polarities 
are not in the 
correct position. 

Reinstall the batteries with their 
negative and positive ends matching 
those indicated in the battery 
compartment.  

The cuff is not 
fastened properly. 

Fasten the cuff correctly. 

Patient moved thier   
arm or body during 
the measurement. 

Make sure the patient remains very 
still during the measurement. 

The cuff position is 
not correct. 

Have the patient sit comfortably still. 
Ensure that the cuff is at the same 
level as the heart.   

 

If the patient has a very weak or 
irregular heart beat, the device may 
have difficulty in determining thier 
blood pressure. 

The unit does   
not take a 
measurement. 
Readings are 
too high or too 
low. 

 

Remove the batteries. Reinstall them 
properly and try the measurement 
again. 

Note:  If the recommendations above do not solve the problem, contact 

your nearest authorized A&D representative. Do not attempt to open 
or repair this product by yourself, as any attempt to do so will render 
your warranty invalid. 

 
 

10. Maintenance 

 

Do not attempt to open the device as the delicate electrical components and 
intricate air unit inside could be damaged. If you cannot solve the problem 
using our troubleshooting guide, request assistance from your authorized 
dealer or from any A&D service group. 

 

The device was designed and manufactured for a long service life. However it 
is generally recommended to have the monitor inspected every 2 years, to 
ensure proper functioning and accuracy. Please contact either your authorized 
dealer or A&D for maintenance. 

 

Содержание UM-101

Страница 1: ...Instruction Manual Manuel d instructions Manual de instrucciones Manuale di Istruzioni Bedienungsanleitung WM PD4001301...

Страница 2: ...uction Manual Contents 1 Dear Customers 2 2 Preliminary Remarks 3 3 Notes for Proper Use 4 4 Parts Identification 6 5 Symbols 7 6 Using the Monitor 8 7 Measurements 9 8 Useful Features 10 9 Troublesho...

Страница 3: ...sphygmomanometer is designed to monitor and display the cuff pressure during cuff inflation and deflation while the healthcare provider determines the patient s blood pressure level by listening for...

Страница 4: ...e mark of conformity accompanied by the reference number to the involved notified body This device is designed for adults only Definitions SYS Systolic Blood Pressure DIA Diastolic Blood Pressure PUL...

Страница 5: ...her liquids If the device is immersed by accident it may require servicing DO NOT use the device before it has endured a full service Do not leave the device in a high temperature or high humidity env...

Страница 6: ...ccessories before any subsequent uses Do not pull or kink the tubes Do not use any organic solvent for example antiseptic solution etc to clean the device Press the ON OFF button after a measurement Y...

Страница 7: ...During measurement After exhausted air Indicates pressure value Indicates pulse rate Display Numerical display Rubber bulb Exhaust valve ON OFF button MARKER button Pressure bar display Cuff 1 5V Batt...

Страница 8: ...are designed to provide special protection against electrical shocks Full Battery The battery power indicator during measurement Low Battery The battery is low when it blinks Replace all batteries wit...

Страница 9: ...ies only The batteries provided with the device are for testing monitor performance and may have a limited life Remove the batteries if the device will not be in use for a prolonged period of time as...

Страница 10: ...lashing the device is ready for measurement If air is remaining in the cuff when the ON OFF button is pressed the display will indicate an error code Switch off the device press the ON OFF button agai...

Страница 11: ...An error message will be displayed if a measurement is taken with insufficient pulses or in a very noisy environment 7 Press the ON OFF button again to turn off the power Note Model UM 101 has an aut...

Страница 12: ...culty in determining thier blood pressure The unit does not take a measurement Readings are too high or too low Remove the batteries Reinstall them properly and try the measurement again Note If the r...

Страница 13: ...Pressure 3 mmHg Pulse 5 Pressure bar display Pressure 4 mmHg Power supply 2 x 1 5 V alkaline batteries LR6 or AA Upper arm circumference 23 33 cm using the medium cuff Classification Type BF EMC IEC...

Страница 14: ...LTD Authorized Representative Established in the European Community Unit 24 26 Blacklands Way Abingdon Business Park Abingdon Oxon OX14 1DY United Kingdom Telephone 44 1235 550420 Fax 44 1235 550485...

Отзывы: