English 10
4. When inflation is complete.
Turn the exhaust valve screw to release air slowly.
Measure the systolic pressure and the diastolic pressure by stethoscopy.
5. The pulse rate is shown on the numerical display when the measurement is
complete, and meets the following conditions.
When you pressurize 80mmHg or higher for the measurement.
When the pressure drops to 20mmHg or lower.
6. Turn the exhaust valve screw counterclockwise to
release all the air from the cuff.
If a measurement is taken with insufficient pressure,
the
mark will be displayed. Re-pressurize the
cuff to a pressure that is about 30 to 40 mmHg higher
than the previous attempt.
An error message
will be displayed if a
measurement is taken with insufficient pulses or in a
very noisy environment.
7. Press the
ON/OFF
button again to turn off the power.
Note: Model UM-101 has an automatic power-off function.
8. Useful Features
Measurement with MARKER Button
You can put a marker at a certain pressure value
when the
MARKER
button is pressed during the
measurement process. Up to 5 markings can be
shown over the range of 40mmHg.
Lifetime Counter
When
the
MARKER
button is pressed while the device
is off, the lifetime counter is displayed.
This counter function indicates the hours the device was
in use and helps to determine when maintenance is
necessary. High digits and Low digits are alternatively
displayed. The example indicates that the device has
been in use for 1,278 hours.
Marker
Marker
High digit
Low digit
Alternatively