English 7
Symbols
Symbols
Function / Meaning
Recommended Action
Appears during measurement when
the cuff is attached loosely
The reading may yield an
incorrect value. Apply the
cuff correctly, and take
another measurement.
Detected rate of IHB/AFib in memory
User
Previous measurements stored in MEMORY.
Average data
FULL BATTERY
The battery power indicator during
measurement.
LOW BATTERY
The battery is low when it blinks.
Replace all batteries with new
ones when the indicator
blinks.
Unstable blood pressure due to
movement during measurement.
Try the measurement
again.
Remain very still during
the measurement.
The systolic and diastolic values are
within 10 mmHg of each other.
Fasten the cuff correctly,
and try the
measurement again.
The pressure value did not increase
during the inflation.
The cuff is not applied correctly.
PUL DISPLAY ERROR
The pulse is not detected correctly.
Blood pressure monitor internal error
Remove the batteries and
press the
START
button,
and then install the batteries
again. If the error still
appears, contact the dealer.
AM
Data taken between 4:00 and 9:59
PM
Data taken between 18:00 and 1:59
x100 [%]
Total number
Number of detected
IHB/AFibs in memory
%IHB/AFib =