E-22
Maintenance
Do not open the device. It uses delicate electrical components and
an intricate air unit that could be damaged. If you cannot fix the
problem using the troubleshooting instructions, contact the authorized
dealer in your area or our customer service department. A&D customer
service can provide technical assistance and spare parts.
Technical Data
Type
UA-651CEL-SM
Measurement method
Oscillometric measurement
Measurement range
Pressure: 0-299 mmHg
Systolic pressure: 60-279 mmHg
Diastolic pressure: 40-200 mmHg
Pulse: 40-180 beats /min
Measurement accuracy
Pressure: ±3 mmHg
Pulse: ±5 %
Power supply
4 x 1.5V batteries (AA) or AC adapter
(TB-233)
(Not included)
Rating
DC6V 3W
Classification
Internally powered ME equipment
(Supplied by batteries)
/Class II
(Supplied by adapter)
Continuous
operation mode
Clinical test
According to ISO81060-2:2013
In the clinical validation study, K5 was
used on 85 subjects for determination of
diastolic blood pressure.
Wireless communication
4G LTE CAT-M1