22
7-5 Calibration Test Using an External Weight
This function tests the balance weighing accuracy using an external mass and outputs the result.
This is available only when the “GLP output (
info
)” parameter is set to “
1
” or “
2
”. (Calibration test
does not perform calibration.)
When the MC-1000 is used
1 Connect the AC adapter and warm up the balance
for at least 30 minutes with nothing on the pan.
2 Press and hold the CAL key until
CCout
is
displayed, then release the key.
3 The balance displays
CC 0
.
z
If you want to change the target weight (a list of
usable weights is shown on page 18), press the
SAMPLE key and proceed to step 4.
z
If you use the target weight value stored in the
balance, proceed to step 5.
4 Specify the target weight value as follows:
SAMPLE key
To switch between the target
weight selection mode (All of
the digits blinking) or the
value adjustment mode (The
selected digits blinking).
With the MC-1000, the digits in
the 2
nd
, 3
rd
, and 4
th
decimal
places can be adjusted.
With the MC-6100, the digits in
the 2
nd
and 3
rd
decimal places
can be adjusted.
RE-ZERO key
To select the target weight
(All of the digits blinking) or
adjust the value (The
selected digits blinking).
Refer to page 18
key
To store the new weight
value. Even if the AC
adapter is removed, the data
is maintained in non-volatile
memory.
CAL key
To cancel the operation and
return to
1
CC 0
.
Press and
hold the key
Release
the key
e.g. Target
weight value
200.0123 g
Select a
weight
To next page
Press