51
11. ID NUMBER AND GLP REPORT
z
The ID number is used to identify the balance when Good Laboratory Practice (GLP) or Good
Manufacturing Practice (GMP) is used.
z
The ID number is maintained in non-volatile memory even if the AC adapter is removed.
z
The output format for GLP/GMP compliant report is selected at “GLP output (
info
)” of the function
table and can be output to a personal computer or printer using the RS-232C serial interface.
z
The GLP/GMP compliant report includes the balance manufacturer, model, serial number, ID
number and space for signature for weight data, and the weight used and results for
calibration or calibration test data.
When the AD-8121B printer is used, the date and time can be printed using the AD-8121B
clock and calendar function. In this case, set the “GLP output (
info
)” parameter to “
1
”.
z
The balance can output the following:
“Calibration report” of the calibration, using the internal mass (Calibration due to changes in
temperature and one-touch calibration.)
“Calibration report” of the calibration, using an external weight.
“Calibration test report” of the calibration test, using an external weight and using the internal mass.
“Title block” and “End block” for the weight data.
11-1 Setting the ID Number
1 Press and hold the SAMPLE key until
ba5fnc
of the function table is displayed.
2 Press the SAMPLE key several times to display
id
.
3 Press the PRINT key. Set the ID number using the following keys.
SAMPLE key
To select the digit to change the value. The selected digit blinks.
RE-ZERO key, MODE key To set the character of the digit selected. Refer to the display
character set shown below.
PRINT key
To store the new ID number and display
ap fnc
CAL key
To cancel the new ID number and display
ap fnc
4
Press the CAL key to return to the weighing mode.
Display character set
11-2 GLP Report
Set the following parameters to output the GLP/GMP compliant report.
z
To print the report, set the “GLP output (
info
)” parameter to “
1
”, set the “Data output pause
(
pU5e
)” parameter to “
1
” and use MODE 3 of the AD-8121B printer. For details on using the
printer, refer to “16-1 Connection to the AD-8121B Printer”.
z
To output the report to a personal computer using the RS-232C serial interface, set the “GLP
output (
info
)” parameter to “
2
”.
Note
If the time and date are not correct, adjust the AD-8121B clock and calendar (HR-A series) or
use “
Cl adj
“ of the function table to adjust the time and date (HR-AZ series).
Содержание HR-250AZ
Страница 86: ...84 MEMO ...
Страница 87: ......