6-2 Smart Range
The smart range for the HR-251AZ/HR-251A consists of the standard range and precision
range (high resolution).
Smart range function
The range switches automatically, depending on the value displayed.
Pressing the RE-ZERO key allows weighing in the precision range, regardless of the tare
value.
The range can be fixed to the standard range by pressing the SAMPLE key.
Precision range and standard range for the HR-251AZ/HR-251A
Unit
Precision range
Standard range
Gram
0.0000 g to 62.0009 g
62.001 g to 252.008 g
Milligram
0.0 mg to 62000.9 mg
62001 mg to 252008 mg
Refer to page 12 for other units.
Example
1 Press the RE-ZERO key to use the precision range.
The balance displays zero.
2 Place a container (tare) on the weighing pan.
When the weight value displayed exceeds the
precision range value, the balance will switch to the
standard range.
3 Press the RE-ZERO key.
The balance displays zero and switches to the
precision range.
4 Place a sample in the container.
When the weight value displayed is within the
precision range value, the balance will perform a
weighing, using the precision range.