36
11. ID Number and GLP Report
The ID number is used to identify the balance when Good Laboratory Practice (GLP) is used.
The ID number is maintained in non-volatile memory even if the AC adapter is removed.
The GLP output format is selected at "GLP output (
info
)" of the function table and can be
output to a personal computer or printer using the RS-232C serial interface.
The GLP output format includes the balance manufacturer, model, serial number, ID
number, date column, time column and a space for signature for the weighing data, the
weight used and results for calibration or calibration test data. When connecting the
AD-8121B printer, the date and time can be printed using the clock function of the
AD-8121B.
The balance can output the following reports for GLP.
"Calibration report" of the calibration, using an external weight.
"Calibration test report" of the calibration test, using an external weight.
"Title block" and "End block" for the weighing data.
11.1.
Setting the ID Number
1 Press and hold the
RANGE
key until
ba5fnc
of the function table is displayed, then
release the key.
2 Press
the
RANGE
key several times to display
id
.
3 Press
the
key. Set the ID number using the following keys.
RE-ZERO
key..... The key to set the character of the digit selected.
Refer to the display character set shown below.
RANGE
key........ The key to select the digit to change the value.
key.......... The key to store the new ID number and display
ba5fnc
.
CAL
key............. The key to cancel the new ID number and display
ba5fnc
.
4 With
ba5fnc
displayed, press the
CAL
key to return to the weighing mode.
Содержание HR-202i
Страница 1: ...Analytical Balance HR 300i HR 202i 1WMPD4000912C ...
Страница 7: ...5 ...
Страница 56: ...54 15 1 External Dimensions Unit mm ...
Страница 61: ...59 PCS Unit of counting mode 16 Unit 10 11 35 Unit Unit 29 WinCT Communication Tools 43 Zero point 57 ...
Страница 62: ...60 MEMO ...