19
7. Response
Adjustment
This function adjusts the response of the balance and reduces the influence of drafts and/or
vibration at the place where the balance is installed. The function has three stages as
follows :
Indicator Parameter Response
Stability
FAST
Cond 0
MID.
Cond 1
SLOW
Cond 2
Fast response, Sensitive value
Slow response, Stable value
1 Press and hold the
MODE
key until
RESPONSE
is
displayed, then release the key.
Press the
MODE
key again quickly.
2 Select a stage of the response adjustment using
the
MODE
key. Either
FAST
,
MID.
or
SLOW
can be selected.
3 The balance automatically returns to the weighing
mode and displays the updated response indicator
for about thirty seconds.
Advise
If the automatic response adjustment is not helpful, specify a parameter for "Condition (
Cond
) "
of "Environment, Display (
ba5fnc
)" with key operation.
.
Response
indicator
Release and wait
Press and hold
Release
Press again
immediately
Select a parameter
Содержание HR-202i
Страница 1: ...Analytical Balance HR 300i HR 202i 1WMPD4000912C ...
Страница 7: ...5 ...
Страница 56: ...54 15 1 External Dimensions Unit mm ...
Страница 61: ...59 PCS Unit of counting mode 16 Unit 10 11 35 Unit Unit 29 WinCT Communication Tools 43 Zero point 57 ...
Страница 62: ...60 MEMO ...