Calibration Instructions
Y
Warm up the scale for more than 5 minutes. Refer to the instruction
manual for how to disable the automatic power-off function, if necessary.
1. Calibration weight
Calibrate the scale with calibration weights equivalent to capacity.
HL-200
i
:
200g ± 0.02g
HL-2000
i
:
2000g ± 0.2g
2. Calibration procedures
(1) Start with the scale turned off. Press the ON:OFF key while pressing
both the RE-ZERO key and UNITS key.
Cal
Cal
Cal
Cal
will be displayed.
(2) Press the RE-ZERO key to calibrate zero.
Cal0
Cal0
Cal0
Cal0
will be displayed.
Make sure there is nothing on the weighing pan. Wait for the stable
indicator to be displayed and press the RE-ZERO key.
Calf
Calf
Calf
Calf
will be displayed in a few seconds.
(3) Place the calibration weight on the center of the weighing pan. Wait for
the stable indicator to be displayed and press the RE-ZERO key.
The display will show
end
end
end
end
.
(4) Remove the calibration weight and the scale will automatically return to
the weighing mode.
(5) Place the calibration weight on the center of the weighing pan again
and check the calibration has been done correctly.
Y
If you press the UNITS key at step (2), a 4 digit number will be
displayed. Press the UNITS key again to return to
Cal
Cal
Cal
Cal
.
End of calibration instructions.