A&D HL-200i Скачать руководство пользователя страница 3

HL-

i

 series Digital Compact Scale 

Calibration Instructions

 

 

Y

 

Warm up the scale for more than 5 minutes. Refer to the instruction 

manual for how to disable the automatic power-off function, if necessary. 

 
1. Calibration weight   

Calibrate the scale with calibration weights equivalent to capacity. 

HL-200

i

 : 

200g ± 0.02g 

HL-2000

i

2000g ± 0.2g 

2. Calibration procedures   

(1) Start with the scale turned off. Press the    ON:OFF    key while pressing 

both the  RE-ZERO  key and  UNITS  key.  

Cal

Cal

Cal

Cal

    will be displayed. 

(2) Press the    RE-ZERO    key to calibrate zero.   

Cal0

Cal0

Cal0

Cal0 

  will be displayed. 

Make sure there is nothing on the weighing pan. Wait for the stable 
indicator to be displayed and press the    RE-ZERO    key. 

 

Calf

Calf

Calf

Calf 

  will be displayed in a few seconds. 

(3) Place the calibration weight on the center of the weighing pan. Wait for 

the stable indicator to be displayed and press the    RE-ZERO    key. 
The display will show   

end

end

end

end 

(4) Remove the calibration weight and the scale will automatically return to 

the weighing mode.

   

(5) Place the calibration weight on the center of the weighing pan again 

and check the calibration has been done correctly. 

Y

 

If you press the    UNITS    key at step (2), a 4 digit number will be 

displayed. Press the    UNITS    key again to return to   

Cal

Cal

Cal

Cal

 . 

 
 

End of calibration instructions. 

 