3-3 Setting the Upper and Lower Limit Values
The results of the comparison are indicated by HI OK LO on the display.
Operating conditions:
?
No comparison
?
Comparison when the weighing data is stable or overloaded, excluding
“near zero”
?
Comparison when the weighing data is stable or overloaded, including
“near zero”
?
Continuous comparison, excluding “near zero”
?
Continuous comparison, including “near zero”
To compare, use:
?
Upper limit value and lower limit value
?
Reference value and tolerance value
Input method:
?
Digital input
?
Weighing input
Note
"Near zero" means within
?
10 digits of the minimum weighing value. For example, using a
GX-2000 in gram mode, near zero is within
?
0.10 g.
Setting example 1
(Continuous comparison, excluding “near zero”, reference value and tolerance value, digital input)
Selecting a comparator mode
1 Press and hold the SAMPLE key until
ba5fnc
of the function table is displayed.
2 Press the SAMPLE key several times to display
Cp fnc
.
3 Press the PRINT key.
4 Press the RE-ZERO key several times to display
Cp 3
.
5 Press the SAMPLE key several times to display
Cp in
.
6 Press the RE-ZERO key several times to display
Cp in 2
.
7 Press the PRINT key to store the selected mode.
Entering the reference and tolerance values
8 With
Cp ref
displayed, press the PRINT key. The current setting is displayed with all the
digits blinking.
?
When the current setting is not to be changed, press the PRINT or CAL key to proceed to step 9.
?
When the current setting is to be changed, press the RE-ZERO key. Change the setting
using the following keys.
SAMPLE key
To select the digit to change the value.
RE-ZERO key
To change the value of the digit selected.
MODE key
To switch the polarity.