23
7.5.
Fixed Voltage Output
The
voltage
output
is fixed under the following conditions:
Step 1 With the display off,
0
V is output (or
0.2
V if the slide switch is set to "0.2V ") when not
weighing during calibration, etc.
Step 2 During
the
re-zeroing
operation in gross full scale output mode (
),
the previous output value is retained.
During the re-zeroing operation in 2
-
digit output mode (
),
0
V is output (or
0.2
V
if the slide switch is set to "0.2V ").
During the re-zeroing operation in 3
-
digit output mode (
),
0
V is output (or
0.2
V
if the slide switch is set to "0.2V ").
Step 3 When
-
E (weighing value is negative overload) is displayed,
0
V is output (or
0.2
V if the slide switch is set to "0.2V ").
Step4 When E (weighing value is positive overload) is displayed,
the following voltages are output.
Example
Voltage output when E is displayed if the slide switch is set to "0V ".
Model
.
、
.
.
、
.
GX-8202M
GX-8202MD
GF-8202M
GF-8202MD
1.000 V
1.025 V
GX-10202M
GF-10202M
1.000 V
1.02 V
GX-12001M
GF-12001M
1.000 V
1.017 V
GX-22001M
GF-22001M
1.000 V
1.009 V
GX-32001M
GX-32001MD
GF-32001M
GF-32001MD
1.000 V
1.006 V